SD3 Series Ultrasonic Pocket Doppler User Manual Troubleshooting
- 46 -
Acoustic output reporting table for IEC60601-2-37
(IEC60601-2-37, Edition 2.1, 2015-06, table 201.103)
Transducer Model: SD3 CW4.0
Working Mode: CW mode
Working Frequency:4.0MHz
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.012
0.13
0.15
N/A
Index component value
0.13
N/A
N/A
0.15
Acoustic
Parameters
p
r.
α
at
z
MI
(MPa)
0.024
P
(mW)
6.69
6.69
N/A
P
1x1
(mW)
N/A
N/A
z
s
(cm)
N/A
z
b
(cm)
2.75
z
MI
(cm)
2.75
z
PII.
α
(cm)
2.75
f
awf
(MHz)
4.00
4.00
4.00
N/A
Other
Information
prr
(Hz)
N/A
srr
(Hz)
N/A
n
pps
N/A
I
pa.
α
at
z
PII.
α
(W/cm
2
)
0.017
I
spta.
α
at
z
PII.
α
or
z
SII.
α
(mW/cm
2
)
16.85
I
spta
at
z
PII
or
z
SII
(mW/cm
2
)
37.72
p
r.
at
z
PII
(MPa) 0.037
Operating
control
conditions
Focus(mm)
Fixed
Depth(mm)
Fixed
Frequency(MHz)
4.00