FTS-3 Fetal Telemetry System Service Manual Installation
- 5 -
WARNING
1
FTS-3 is provided for the use of qualified physicians or personnel professionally
trained. They should be familiar with the contents of this user manual before
operation.
2
FTS-3 is not intended for use in intensive care units (ICU), operating rooms or for
home use.
3
Do not switch on the base station until all cables have been properly connected and
verified.
4
To avoid the danger of introducing additional hazards, do not perform any
unauthorized modification of FTS-3.
5 EXPLOSION HAZARD
- Do not use FTS-3 in the presence of flammable anesthetics
or other materials.
6 SHOCK HAZARD
- Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power cord.
7
Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously. Do not touch the signal input or output connector and the patient
simultaneously.
8
The service personnel should be familiar with the operation of FTS-3. Refer to
F9+F9E Fetal & Maternal Monitor User Manual
for details.
9
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all
configurations shall comply with the valid version of the system standard IEC/EN
60601-1. Anybody who connects additional equipment to the signal input connector or
signal output connector to configure a medical system must ensure that the system
complies with the requirements of the valid version of the system standard IEC/EN
60601-1. If in doubt, consult our technical service department or your local distributor.
10
Do not exceed the maximum permitted load when using multiple portable
socket-outlets to supply the system.
11
SHOCK HAZARD -
Don’t connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer. If multiple instruments are connected to a patient, the sum of the
leakage currents may exceed the limits given in the IEC/EN 60601-1 and may pose a
safety hazard. Consult your service personnel.
12
Do not use the additional multiple portable socket-outlet or extension cord in the
medical electrical system, unless it’s specified as part of the system by manufacturer.
And the multiple portable socket-outlets provided with the system shall only be used
for supplying power to equipment which is intended to form part of the system.
13
Multiple portable socket-outlets shall not be placed on the floor.
14
SHOCK HAZARD
- Do not remove the top panel cover during operation or while
power is connected.