VE-100 Veterinary Electrocardiograph User Manual
-22-
Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align lead wires of patient cable to avoid twisting;
3)
Clean electrode area on a short distance with alcohol;
4)
Daub the electrode area with gel evenly;
5)
Place a small amount of gel on the metal part of electrode clamp;
6)
Clamp the electrode to the electrode area. Attach all electrodes in the same way.
WARNING
:
♦
Be sure that all electrodes have been connected to the animal correctly before
operation.
♦
Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrode, do not come in contact with earth or any other
conducting objects.
♦
There is no danger when using the electrocardiograph with electrical
stimulation equipment. However, the stimulation units should only be used at
a sufficient distance from the electrodes .If in doubt, the animal should be
disconnected from the device.
♦
Electrodes with defibrillation protection should be used while defibrillating.
♦
Do not touch the unit casing during defibrillation.
4.5 Inspection before Switching on
In order to avoid safety hazards and get good ECG records, the following inspection
procedures are recommended before turning on the unit and beginning operation.
1)
Environment
:
♦
Check and make sure that there is no electromagnetic interference source around the
equipment, especially large medical electrical equipment such as electrosurgical
equipment, radiological equipment and magnetic resonance imaging equipment etc.
Switch off these devices when necessary.
♦
Keep the examination room warm to avoid muscle action voltages in ECG signal
caused by cold.