EPX-SERISE USER MANUAL Rev. G
18
4.2.5 EPX-F4000
IDENTIFICATION LABEL-HIGH TENSION TANK
This is label is attached on side of the HV mono block tank inside of the diagnostic x-ray system.
(Example)
Output power: 4.0kW , KVP range: 50-110kVp
mA range: 20-80mA,
Line Power: 200-240V , 50/60Hz
Obelis S.A
Avenue de Tervuren, 34, bte 44 1040 Brussels, Belgium / E-mail:[email protected]
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03
Ecotron Co., Ltd
Made by Korea
HV Mono Block Tank
#504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu,
Seoul, Korea
EMB-F4000 Series
Inserted X-ray Tube
Focal spot: 1.8mm x 1.8mm, Target angle: 15 degrees
Anode Heat Storage: 42kHU
Inherent Filtration: At least 0.5mmAl
OX/110-15
Certification Statement
▪ This product complies with 21 CFR 1020.30
–
1020.31
▪ This product complies with 21 CFR Subchapter J.
REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM
This label is attached to rear cover of the diagnostic x-ray system. (Example)
Input Power: 200-240V , 50/60Hz, 11kVA, Standby 150VA
Output power: 4.0kW @ 50kV, 80mA
KVP range: 50-110kVp, mA range: 20-80mA, mAs Range: 0.4-100mAs
▪ Total Filtration: 3.0mmAl
(Tube: 0.5mmAl, Fixed Added Filter: 2.0mmAl, Collimator: 0.5mmAl)
▪ Time interval between Max kV and Max mAs exposure: rest time 5min
Obelis S.A
Avenue de Tervuren, 34, bte 44 1040 Brussels, Belgium / E-mail:[email protected]
Representative: Mr.Gideon ELKAYAM / TEL:32.2.732.59.54, FAX:32.2.732.60.03
Ecotron Co., Ltd
Made by Korea
Diagnostic X-ray System
#504~505Ho, Hanshin IT Tower II, 47, Digital-ro 9-gil, Geumcheon-Gu, Seoul, Korea
EPX-F4000 (S/W: Ver )
Inserted X-ray Tube
▪ Focal Spot: 1.8mm x 1.8mm, Inherent Filtration: At least 0.5mmAl
Target Angle: 15 degrees, Anode Heat Storage: 42kHU
OX/110-15
2015.03
Certification Statement
▪ This product complies with 21 CFR Subchapter J.
CAUTION
▪ Only an experienced expert should operate the unit.
▪ Do not use with wet hands.