50
Note
. A zero value of the threshold level makes the alarm inactive.
To save the measurement result of photon-ionizing radiation DER in the
nonvolatile memory, press and hold the THRESHOLD button in the measurement
mode until the “Arch” symbols are displayed on the LCD (Figure 5). Then release
the THRESHOLD button. Switching to the submode of measurement result saving
should be confirmed by a short press of the MODE button. Press shortly the
THRESHOLD button to cancel the action. If the buttons are not pressed for 30 s, the
radiometer automatically returns to the mode of photon-ionizing radiation DER
measurement.
If the “FULL” symbols are displayed on the LCD of the radiometer instead of
the “Arch” symbols (Figure 10), there is no free space in the nonvolatile memory of
the radiometer, and, correspondingly, the next measurement results cannot be saved.
Содержание STORA-TU RKS-01
Страница 1: ...RKS 01 STORA TU GAMMA BETA RADIOMETER DOSIMETER Operating manual...
Страница 2: ......
Страница 20: ...18 Figure 1 External view of the radiometer top view...
Страница 22: ...20 Figure 2 External view of the radiometer bottom view...
Страница 38: ...36 Figure 3 LCD of the radiometer the mode of photon ionizing radiation DER measurement...
Страница 42: ...40 Figure 4 LCD of the radiometer viewing the value of specified statistical error...
Страница 46: ...44 Figure 7 LCD of the radiometer the submode of alarm threshold level programming...
Страница 55: ...53 Figure 11 LCD of the radiometer the submode of measurement result saving of photon ionizing radiation DER...
Страница 59: ...57 Figure 12 LCD of the radiometer the mode of surface beta particles flux density measurement...
Страница 63: ...61 Figure 13 LCD of the radiometer viewing the value of specified statistical error...
Страница 67: ...65 Figure 16 LCD of the radiometer the submode of alarm threshold level programming...
Страница 73: ...71 Figure 19 LCD of the radiometer the clock mode...
Страница 76: ...74 Figure 21 LCD of the radiometer the submode of time and date correction year programming...
Страница 78: ...76 Figure 22 LCD of the radiometer the submode of time and date correction date programming...
Страница 81: ...79 Figure 23 LCD of the radiometer the alarm clock mode...
Страница 82: ...80 Figure 24 LCD of the radiometer the submode of alarm clock activation time programming...
Страница 87: ...85 Figure 26 LCD of the radiometer the mode of data communications control with the PC...
Страница 89: ...87 Figure 27 LCD of the radiometer the mode of data communications control with the PC...
Страница 92: ...90 Figure 28 LCD of the radiometer the mode of measurement results viewing stored in the nonvolatile memory...
Страница 94: ...92 Figure 29 LCD of the radiometer measurement results viewing...
Страница 97: ...95 Figure 30 LCD of the radiometer the submode of measurement results clearing stored in the nonvolatile memory...
Страница 121: ...119 APPENDIX Figure A 1 Plot of specified statistical error versus photon ionizing radiation DER...
Страница 122: ...120 APPENDIX Figure A 2 Plot of specified statistical error versus surface beta particles flux density...