Installation
Page 31
2.
Installation
2.1
Initial Check
Packaging
The Serinus 40 is transported in packaging which is specifically designed to minimise the effects of
shock and vibration during transportation. Ecotech recommends that the packaging be kept if there
is a likelihood that the instrument is going to be relocated.
Note:
The red plastic caps that seal the pneumatic connections during transport must be removed
prior to operation.
Opening the Instrument
To check the interior of the instrument:
1.
Undo the screws located in the rear panel.
2.
Open the chassis lid by releasing the latch (pressing the button) located on the front panel in the
top left-hand corner, then slide the lid backwards.
3.
To completely remove the lid, slide the lid backwards until the rollers line up with the gaps in the
track and pull the lid upwards to remove from the instrument (refer to Figure 3).
4.
Check that all pneumatic and electrical connectors are connected.
5.
Check for any visible and obvious damage. If damage exists contact your supplier and follow the
instructions in (Claims for Damaged Shipments) and (Shipping Discrepancies) at the front of this
manual.
Figure 3 – Opening the instrument
Содержание Serinus 40
Страница 1: ...Introduction Page 1 Serinus 40 Oxides of Nitrogen Analyser User Manual Version 2 2 www ecotech com ...
Страница 2: ...Serinus 40 User Manual 2 2 Page 2 This page is intentionally blank ...
Страница 18: ...Serinus 40 User Manual 2 2 Page 18 This page is intentionally blank ...
Страница 30: ...Serinus 40 User Manual 2 2 Page 30 This page is intentionally blank ...
Страница 62: ...Serinus 40 User Manual 2 2 Page 62 This page is intentionally blank ...
Страница 84: ...Serinus 40 User Manual 2 2 Page 84 This page is intentionally blank ...
Страница 114: ...Serinus 40 User Manual 2 2 Page 114 This page is intentionally blank ...
Страница 118: ...Serinus 40 User Manual 2 2 Page 118 7 1 Flow Fault Figure 67 Zero flow fault troubleshooting flow chart ...
Страница 119: ...Troubleshooting Page 119 7 2 Noisy Unstable Readings Figure 68 Noisy or unstable readings troubleshooting flow chart ...
Страница 138: ...Serinus 40 User Manual 2 2 Page 138 This page is intentionally blank ...
Страница 146: ...Serinus 40 User Manual 2 2 Page 146 This page is intentionally blank ...