4.4.2
Analog Inputs
The instrument is also equipped with three analog inputs with resolution of 15 bits plus polarity,
accepting a voltage between 0 - 5 V. These go directly to the microprocessor and should be protected
to ensure static/high voltage does damage the main controller PCA (instrument warranty does not
cover damage from external inputs).
4.4.3
Digital Status Inputs
The instrument is equipped with eight logic level inputs for the external control of the instrument such
as Zero or Span sequences. Each input has a terminating resistor which can be either PULL UP or PULL
DOWN. This is set using the jumper JP1 on the rear panel PCA (refer to Figure 85).
4.4.4
Digital Status Outputs
The instrument is equipped with eight open collector outputs which will convey instrument status
condition warning alarms such as no flow, sample mode, etc. Two of the digital outputs can be set so
that there is +5 V and +12 V available on the 25 pin connector for control purposes, instead of digital
outputs 0 and 1.
In the default jumper locations (refer to Figure 85) these two outputs will function normally as open
collector outputs. If moved to the position closer to the 25 pin connector, then the DO 0 will supply
+12 V and DO 1 will 5 V.
The +12 V and +5 V supplies are limited to about 100 mA each.
Each digital output is limited to a maximum of 400 mA. The total combined currents should not exceed
2 A.
Figure 85
–
25 Pin Rear Panel PCA (Default Jumpers Highlighted)
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...