T-Surface Sensor                                        

                                                   

 

 

 

 

Installation 

1.  Fix  the  sensor  cable  at  the  measurement 

object (e.g.  stem  or  branch),  by  using  the 
included  elastic  and  UV-resistant  rubber 
band. 

2.  Place  and  fix  the  sensor  head  at  the  de-

sired measurement position. In case stem 
surface  temperature  is  to  be  measured, 
this can also be done by using the includ-
ed elastic and UV-resistant rubber band.           

3.  Leave  about  30  cm  of  additional  cable 

slackly  hanging  between  the  two  above  described  fixation  points,  so  that  the  sensor 
head is free of any pulling force. 

4.  Ensure  a  good  thermal  contact  between  surface  of  the  measurement  object  and  the 

head of the temperature probe 

5.  Strain relieve the sensor cable leading to the data logger as close as possible to the in-

strumented measurement object, e.g. on a peg firmly hammered into the ground. 

You need further assistance?  

In case you need further assistance for installation, please do not hesitate to contact us.  

 

4. Wiring and Logger Configuration 

The T-Surface sensor is compatible with our DL 18 data logger (ordered with stereo plug), 
as  well  as  with  a  wide  range  of  other  available  data  loggers,  e.g.  CR1000  (ordered  with 
bare cable ends). However, note that suitable loggers have to provide a precise and sta-
ble,  switched (sensor  should  only  be  powered  100ms  before  and  during  measurements) 
excitation voltage (V

ex

) of usually 2500 mV.  

Connecting the sensor with bare cable ends, e.g. to a CR1000 logger: 
The  sensor  can  be  connected  either  in  differential  (wiring  diagram  a),  or  in  single-ended 
(wiring diagram b) mode. 

 

Figure A: Differential wiring 

 
 

 

Figure B: Single-ended wiring

 

 
 
 