44
ecom-ST 1.0
Technical Data
EC = Electrochemical sensor
IR = Infrared sensor
K = Catalytic Sensor
T = Thermocouple NiCr-Ni
PT = PT 2000
D = DMS bridge
MV* = Measured value (higher value prevails)
Measured values
Range
Accuracy
O
2
EC 0 vol. %
−
21 vol. %
±0,3 vol. %
CO
EC 0 ppm
−
4000 ppm
±20 ppm / 5 % MV*
CO H
2
- comp.
EC 0 ppm
−
10.000 ppm
±20 ppm / 5 % MV*
CO%
EC 0 ppm
−
63.000 ppm
±100 ppm / 5 % MV*
NO
EC 0 ppm
−
5000 ppm
±5 ppm / 5 % MV*
NO ExtraLow
EC 0 ppm
−
300 ppm
±2 ppm / 5 % MV*
NO
2
EC 0 ppm
−
1000 ppm
±5 ppm / 5 % MV*
NO
2
Low
EC 0 ppm
−
100 ppm
±5 ppm / 5 % MV*
SO
2
EC 0 ppm
−
5000 ppm
±10 ppm / 5 % MV*
SO
2
Low
EC 0 ppm
−
100 ppm
±5 ppm / 5 % MV*
H
2
EC 0 ppm
−
20.000 ppm
±100 ppm / 5 % MV*
H
2
S
EC
±10 ppm / 5 % MV*
CO
2
IR 0 vol.%
−
20 vol.%
±0,5 vol. % / 5 % MV*
CO
2
IR 0 vol.%
−
100 vol.%
±5 vol. %
CH
4
IR 0 vol.%
−
5 vol.%
±0,2 vol. % / 5 % MV*
C
x
H
y
K
0 vol.%
−
4 vol.%
T-Gas
T
0 °C
−
500 °C
±2 °C / 1,5 % MV*
T-Air
PT 0 °C
−
99 °C
±1 °C
pressure
D
0 hPa
−
± 100 hPa
±0,5 hPa / 1 % MV*
Air pressure
D
300 hPa
−
1100 hPa
±2 hPa