Patient Safeguard System Installation Tool Directions for Use                                                          Ecolab, Inc. 

Page 13 

have 

worked  correctly).    The  Tool  will  then  display  the  Beacon’s  address

the  Beacon’s  battery  level  as  a 

percentage, 

and the Beacon’s firmware revision.  If the radios are active, the Tool’s display may be too dim to 

see.    To  correct  this  press  the  Select  key  and  the  Tool  will  turn  off  the  radios  and  turn  the  display  on  for  5 
seconds. 

5.11.2 Bentley Mode 

 

 

Figure 10 

 Display in Bentley Mode, System B 

Bentley  mode  is  used  to  test  the  433  MHZ  radio  that  is  used  by  a  Beacon  to  communicate  with  the  Bentley 
Network.    The  Tool  will  only  respond  five  times  to  Beacons  in  this  mode,  and  then  the  Tool  will  automatically 
switch to  Beacon mode.  This is done to prevent the Tool from flooding the network with extra messages.  The 
dashboard 

will have to be checked to verify that the Beacon’s message has worked its way through 

the network 

successfully.  The d

ashboard is checked by using the “Search” function and searching for the Beacon’s address

or searching for a Badge address of 0A000001.  

If the radios are active, the Tool’s display may be too dim to s

ee.  

To correct this press the Select key and the Tool will turn off the radios and turn the display on for 5 seconds. 

5.11.3 Range Adjust Mode 

 

 

Figure 11 

 Display in Range Adjust Mode, System B 

Range Adjust mode is used to change the range settings of the Beacons.  The Tool will display the current range 
setting of the last Beacon it communicated with.  If the radios are active, 

the Tool’s display may be too dim to see.  

To correct this press the Select key and the Tool will turn off the radios and turn the display on for 5 seconds.  If 
either the Up key or the Down key is pressed the Tool will then lock onto the last Beacon it communicated with.  
The Up key and Down key are then used to change the range setting of the Beacon, when the desired range 
setting is displayed, pressing the Select key will cause the Tool to transmit this setting to the Beacon it has locked 
onto, when the communication process is successfully completed the Yellow and Red LEDs will flash.  If any 
other Beacon is in range the tool will not send the command for it to change its range setting.  If the Tool has 
locked onto the wrong Beacon or if it is not desired to change the range setting of a Beacon the Tool can be 
unlocked by pressing the Menu key and reselecting Range Adjust mode and no changes will be made. 

5.11.4 Bed/Stretcher Mode 

The Bed/Stretcher mode is used to change the part of a Bed Beacon’s configu

ration that requires the Beacon to 

check for a power detect signal from the AC Sensor (FAS1542) before the Bed Beacon transmits to any Badges 

present.  The Tool first must communicate with a Bed Beacon to get the Beacon’s current Bed/Stretcher setting.  

If 

the radios are active, the Tool’s display may be too dim to see.  To correct this press the Select key and the 

Tool will turn off the radios and turn the display on for 5 seconds.  If either the Up key or the Down key is pressed 
the Tool will then lock onto the last Bed Beacon it communicated with.  The Up and Down keys are then used to 
toggle the setting.  Pressing the Select key will cause the Tool to transmit this setting to the Bed Beacon it has 
locked onto, when the communication process is successfully completed the Yellow and Red LEDs will flash.  If 
any other Bed Beacon is in range the tool will not send the command for it to change its Bed/Stretcher setting.  If 
the Tool has locked onto the wrong Bed Beacon or if it is not desired to change the Bed/Stretcher setting of a Bed 
Beacon the Tool can be unlocked by pressing the Menu key and reselecting Bed/Stretcher mode and no changes 
will be made. 

Beacon Address 

Tool Mode 

Beacon Address 

Tool Mode 

Beacon Range Setting 

Содержание FAS1543

Страница 1: ...1 961 5585 Fax 561 961 5587 Certification Exhibit FCC ID Z9O FAS1543 02 IC 10060A FAS154302 FCC Rule Part 15 209 15 231 15 249 ISED Canada Radio Standards Specification RSS 210 ACS Project 16 2065 Man...

Страница 2: ...Patient Safeguard System Installation Tool Directions for Use For Use With FAS1543 PSS Installation Tool DOC8022 Revision A...

Страница 3: ...l 6 4 0 Operation with System A 7 4 1 Boot Post 7 4 2 Changing Modes 7 4 3 Turning the Display Back On 7 4 4 Beacon Communications Testing 7 4 5 Testing a Beacon s Bentley Communications 7 4 6 Testing...

Страница 4: ...dashboard 15 7 0 Cleaning the Tool 15 Appendix A Certification and Safety Approvals 16 FCC Statement 16 Industry Canada 16 Appendix B Warranty 17 Table of Figures Figure 1 FAS1543 Beacon Installation...

Страница 5: ...stallation DOC1054 Patient Safeguard System SmartLink Dispenser Beacon Module Installation Maintenance and Repair Manual DOC8043 Patient Safeguard System Dashboard Directions for Use Abstract The Func...

Страница 6: ...n the dispenser is activated dispense event collect dispense event data from that Badge and then transmit the collected data to a network where it will ultimately be sent to an offsite server for proc...

Страница 7: ...Patient Safeguard System Installation Tool Directions for Use Ecolab Inc Page 5 System A System B...

Страница 8: ...s Power Switch The Power Switch is used to turn the Tool on and off when using battery power Peripheral Connector The Peripheral Connector is used to connect either an external Antenna FAS1540 or an A...

Страница 9: ...a Beacon and a Badge can be tested with the Tool the Tool emulates a Badge The Tool is placed in Beacon mode and the three indicator LEDs give an indication of the quality of the communications at the...

Страница 10: ...ge and enter the address 1E000001 under the device and select PSS Events from the event list Click the Search button and the dashboard should display an event that has an ET Event Time that correspond...

Страница 11: ...and the Tool will turn off the radios and turn the display on for 5 seconds 4 9 3 Range Adjust Mode Range Adjust mode does not work with System A 4 9 4 Bed Stretcher Mode The Bed Stretcher mode does n...

Страница 12: ...section 4 9 6 7 Network mode see section 4 9 7 8 Tool Info mode see section 4 9 9 4 9 9 Tool Info Mode Figure 7 Display in Tool Info Mode First Screen System A Figure 8 Display in Tool Info Mode Secon...

Страница 13: ...the display on for a brief period of time so the results can be read 5 5 Testing a Beacon s Bentley Communications The Tool can be used to check the operation of a Beacon s Bentley radio by placing th...

Страница 14: ...h button and the dashboard should display an event that has an ET Event Time that corresponds to when the Tool s Select key was pressed From the dashboard the relative strength of the Tool s signal ca...

Страница 15: ...key and Down key are then used to change the range setting of the Beacon when the desired range setting is displayed pressing the Select key will cause the Tool to transmit this setting to the Beacon...

Страница 16: ...en the Tool and Badge should be at least 40 inches when the Badge passes the Tool s test Figure 14 Badge Mode Display System B 5 11 7 Network Mode Network mode is used to test the integrity of the Ben...

Страница 17: ...shooting 6 1 The Tool s display will not turn on when the Select key is pressed A Ensure that the Power Switch is pressed down on the right side B Replace the two AA alkaline batteries 6 2 Beacon even...

Страница 18: ...the equivalent isotropically radiated power e i r p is not more than that necessary for successful communication Conform ment la r glementation d Industrie Canada le pr sent metteur radio peut fonctio...

Страница 19: ...the following cases a Loss or damage to the equipment due to abuse mishandling accident or failure to follow installation or use instructions b If the equipment is defective as a result of leaking bat...

Отзывы: