background image

  

ECO SENSORS, INC.

OZONE ANALYZER

Model UV-100

INSTRUCTIONS FOR USE

GENERAL

The model UV-100 is a UV absorption analyzer
designed specifically for ozone. It uses a mercury
lamp filtered for absorption at 254 nm as is typical of
other commercial instruments. The Eco Sensors, Inc.,
design is very compact and uses the latest digital
microprocessor technology. The UV-100 covers three
ranges: 0-1, 0-10 and 0-999 PPM. Its outputs are a
LCD digital display, 0-2.5 VDC, 4-20 mA, and a time
and date stamped digital data stream for external
readout and data logging. The data output are via
serial port and USB. There is also an internal data
logger which can store and forward approximately 36
hours of data. The theory of operation is described in
Appendix D.

Panel LEDs indicate "on", low lamp output, low flow rate, and the ozone concentration at
the sampling point exceeds 0.1 PPM.

The instrument is conditionally warranted for one year. Read the warranty statement at 
the end of this manual.

OPERATION

Plug the AC adapter into AC power and plug the adapter's output plug into the power 
jack on the instrument. Find the power switch on the rear panel and turn the instrument 
on. The green LED "on" indicator and the LCD display backlighting should illuminate. 

Allow at least 1/2 hour warm-up. 

More than 1 hour is preferred if time permits. Initial 

PPM readings may start out high and possibly drop to zero for a few minutes. Then the 
readings will tend towards the actual value as the systems settles into equilibrium.

The compact, high-speed sample pump in the instrument makes a buzzing sound. 
This is normal. 

1

Содержание UV-100

Страница 1: ... The theory of operation is described in Appendix D Panel LEDs indicate on low lamp output low flow rate and the ozone concentration at the sampling point exceeds 0 1 PPM The instrument is conditionally warranted for one year Read the warranty statement at the end of this manual OPERATION Plug the AC adapter into AC power and plug the adapter s output plug into the power jack on the instrument Fin...

Страница 2: ... unless it is set to a preferred range by the Function control on the front panel See further instructions on Range Selection below Front Panel Back Panel Use only the sample air filter supplied with the instrument Other filters especially those with a greater pressure drop can materially affect the readings This will be noticed if the instrument s zero is significantly above 0 ppm with no ozone p...

Страница 3: ...ion MENU SECTIONS Upon start up the UV 100 will display software version number then date and time Pressing the function knob will open the main menu Scale SCL Menu Pressing the function knob again enters the scale menu The UV 100 independently controls the 0 2 5 VDC and 4 20mA scales This feature is very useful in multi level control applications See the Application Examples section for more info...

Страница 4: ...download END command stops storage of measurements Data will be kept in memory until an Overwrite command is given Output To connect the UV 100 to the computer refer to Connections To External Equipment section To download connect the UV 100 to a computer and select XMT The UV 100 will output all measurement records in memory in a comma delimited serial stream Output Examples Normal Mode 0 00 298 ...

Страница 5: ...5 2008 7 0 03 298 4 575 8 09 40 29 27 05 2008 8 0 03 298 4 575 8 09 40 39 27 05 2008 Sample Number Configuration CFG Menu Use to set the relay set points and date and time Relay The relay set points are used to control a variety of external equipment such as Ozone generators circulation fans indicator lights and alarms CAUTION The relay is rated at 240V and 5A For controlling large motors and othe...

Страница 6: ...sure at sea level A convenient formula for approximating the correction for Ozone analyzers other than the UV 100 is For every 1 000 meters of altitude add 14 to the uncorrected reading For every 1 000 feet of altitude add 5 to the uncorrected reading PANEL LED INDICATORS 0 1 PPM Reading is above the OSHA Limit for 8 hour exposure to Ozone Low Flow Input sample draw is less than the average of 1 L...

Страница 7: ...ms in use if you have problems installing or using the drivers on our disc contact Tech Support at Eco Sensors tech ecosensors com Set Point Relay Contacts A wiring plug is provided to enable component control via relay Refer to Appendix X for application examples using this relay SERVICE AND MAINTENANCE There are no user serviceable parts in this instrument Opening the enclosure will void the war...

Страница 8: ...ministered Call a qualified electrician if you have any doubts about voltages currents electrical practice etc While not necessary for typical uses for highest precision results in electrically noisy locations the instrument should be grounded to earth to achieve maximum stability WARRANTY This product is warranted against defects in materials and workmanship for one year following the date of pur...

Страница 9: ...FOR YOUR RECORDS Eco Sensors is a registered trademark of Eco Sensors Inc Eco Sensors Inc 2004 Revision 08 11 For brochures application and tech notes and other useful information visit our extensive website at www ecosensors com E mail us at sales ecosensors com APPENDIX A RS232 SERIAL DATA SETUP AND READOUT EXCLUDING WINDOWS 7 1 Open HYPERTERMINAL from the ACCESSORIES folder of the PROGRAMS dire...

Страница 10: ...NECTION DESCRIPTION Click OK 3 Choose which port you will use to connect to the computer The default serial COM port on most PC computers is COM1 Click OK 4 Enter PORT SETTINGS as above Bits per second 9600 data rate Data bits 8 10 ...

Страница 11: ...l can be transferred to Windows 7 Copy hypertrm exe from Program Files Windows NT directory and hypertrm dll from the Windows System32 directory of the Windows XP disk to APPENDIX B UV 100 USB DIRECT CONNECT PROCEDURE Items UV 100 USB Cable PC Computer with Windows 2000 Vista or XP USB to UART Driver Disk 1 Install USB to UART Driver Disk in the computer s CD ROM drive 2 With the UV 100 off attach...

Страница 12: ...7 Run the Installation wizard 8 Finish the driver installation 12 ...

Страница 13: ...imilar Power relay s for generator and fan optional Indicator light for OSHA limit 4 20 mA current loop To destroy mold the space is held at a very high Ozone level for a predetermined amount of time The generator is controlled using the set point relay on the back of the UV 100 to hold Ozone After the Ozone dwell interval is complete generator power is shut off or UV 100 generator ON level is set...

Страница 14: ...inlet length and loss from the moisture trap LONG TERM DATA LOGGING EXAMPLE The internal memory for data storage for the UV 100 is limited to approximately 36 hours For long term logging the best method is voltage data logging using the Eco Sensors DL 3 and the 0 2 5 VDC output on the 9 pin connector Items UV 100 Computer Eco Sensors DL 3 Voltage Data Logger D9 cable with pin breakout HOBO Data Lo...

Страница 15: ...th high precision and accuracy The ozone molecule has an absorption maximum at 254 nm coincident with the principal emission wavelength of a low pressure mercury lamp Fortunately few molecules found at significant concentrations in the atmosphere absorb at this wavelength However interferences such as organic compounds containing aromatic rings can occur in highly polluted air Figure 1 is a schema...

Страница 16: ...y UV absorption requires no external calibration it is an absolute method However non linearity of the photodiode response and electronics can result in a small measurement error Therefore each instrument is compared with a NIST traceable standard ozone spectrophotometer in the laboratory over a wide range of ozone mixing ratios These results are used to calibrate the Ozone Monitor with respect to...

Страница 17: ...00 Hi Efficiency Sample Line Filter and Housing PN FLTRS LARGE 126 00 Hi Efficiency Sample Line Filter Elements 10pcs PN FILTERPAK LARG 63 00 Heavy Duty External Pump replaces internal pump PN EXT PUMP HD 440 00 C Type Power Plug 240 VAC 50Hz Adapter 75 00 17 ...

Отзывы: