TB9: SENSOR POWER:

 Refer to Appendix A for an illustration and more details. Connect power conductor (red) from either:  

+5 V out to SM-module TB1 +5 V   

OR   

+12-24 V out to SM-module TB2 +12-24 V   

WARNING: 

Do not connect ‘+12-24 V Out’ terminal on the OS-6 to the SM-module TB1 ‘+5 V’.  This will damage the SM-module 

and void the warranty.  

 

 

Power Connection 

Disconnect power before opening the case to access the OS-6. Use only one of the following two methods to power the OS-6: 

1. 

AC Adapter (included): 12-24 VDC, 300 mA (minimum), barrel connector. Refer to Tech Note P-101 at 
www.ecosensors.com/support for additional AC adapter specifications. 

2. 

TB1: 12-24 VDC, 300 mA (minimum), wired connection (16-26 gauge) 

 

 

Initial Operation & Warm-Up

 

All standard connections between the OS-6 and the SM-module are made at the factory. Once power is connected to the OS-6 using one of 
the methods described above, the green power/data LED will illuminate and then blink at a 1-second interval as data is received from the SM-
module. The LCD will display digits. 

Power the OS-6 and connected SM-module for 1-hour before response testing. Prior to on-site process integration, operate on power for 24 
hours. We recommend testing the instrument for positive response with an ozone generator when the instrument is received and again at the 
site where the instrument is installed. 

 

 

Set Point and Alarm Condition Settings

 

Five  push-buttons  are  mounted  on  the  circuit  board 
behind the  wiring compartment cover. Remove the cover 
(two  screws)  and  locate  the  buttons  in  the  upper-left 
corner.    From  left  to  right,  they  are:  FAIL,  MAX,  MIN, 
INCR and DECR.  

FAIL

:  Time  (in  minutes)  that  the  ozone  level  can  remain 

above  MAX  set  point  or  below  MIN  set  point  before  the 
'‘Unable to Control Ozone” indicator and relay engage. To 
disable the “Unable to Control Ozone” indicator and relay 
(TB6), set FAIL to 0999 (available on instruments shipped 
after 2012-10-15, call us for details).  

MAX

:  Maximum ozone set point at which the generator is 

turned off.  

MIN

:  Minimum ozone set point at which the generator is turned on.  

INCR

:  Increase set point value.  

DECR

:  Decrease set point value.  

Change a set point by pressing FAIL, MAX, or MIN. The current value of that set point will appear on the display. Press INCR or DECR to 
adjust the value. MIN will not adjust higher than MAX.  

 

 

Use either 

FAIL   MAX    MIN       INCR DECR

Содержание OS-6

Страница 1: ...variety of other important ozone control parameters x Set points for ozone generator control relays are digitally controlled with numeric set points This makes precise ozone control easy to set up eve...

Страница 2: ...power analog outputs relay outputs and Monitor Mode selector are located in the Wiring Compartment For access disconnect the OS 6 from power and remove the two screws and cover The connection board w...

Страница 3: ...he LCD will display digits Power the OS 6 and connected SM module for 1 hour before response testing Prior to on site process integration operate on power for 24 hours We recommend testing the instrum...

Страница 4: ...ached SM module to determine the measurement range The OS 6 is shipped from the factory with the Analog Output Full Scale Range set to match the installed SM module For further details refer to the sp...

Страница 5: ...defects in materials and workmanship for one year following the date of purchase by the original owner This warranty does not include damage to the product that results from misuse accident dropping m...

Страница 6: ...6 Wiring Compartment Panel SM Module User Replaceable Sensor Connects Here SM Module OS 6 Wiring Compartment Panel Power In GND 12 24 VDC 3 Wire Cable Power Out Data In GND 5VDC WARNING Never connect...

Страница 7: ...al SM module Connection A power relay is required if the load exceeds the OS 6 relay ratings Power Relay Analog Outputs to Process Controller Refer to SM Module Wiring Illustrations N O COM N C Genera...

Отзывы: