55
57-646 ECLIPSE Guided Wave Radar Transmitter - F
OUNDATION
fieldbus
™
6.2 agency Specifications – Explosion Proof installation
Factory Sealed: this product has been approved by Factory Mutual Research (FM) as a Factory Sealed device.
notE:
Factory Sealed: No Explosion Proof conduit fitting (EY seal) is required within 18" of the transmitter. However, an
Explosion Proof conduit fitting (EY seal) is required between the hazardous and safe areas.
6.0 Reference Information
6.1 Agency Approvals
These units are in compliance with the EMC-directive 2014/30/EU,
the PED-directive 2014/68/EU and the ATEX directive 2014/34/EU.
Explosion Proof (with intrinsically Safe Probe)
US/Canada:
Class I, Div 1, Group B, C and D, T4
Class I, Zone 1 AEx d/ia [ia IIC Ga] IIB + H2 T4 Gb/Ga
Class I, Zone 1 Ex d/ia [ia IIC Ga] IIB + H2 T4 Gb/Ga
Ta = -40ºC to +70ºC
Type 4X, IP67
Flame Proof
ATEX – FM14ATEX0041X:
II 2/1 G Ex d/ia [ia IIC Ga] IIB + H2 T6 to T1 Gb/Ga
Ta = -40ºC to +70ºC
IP67
IEC- IECEx FMG 14.0018X:
Ex d/ia [ia IIC Ga] IIB + H2 T6 to T1 Gb/Ga
Ta = -40ºC to +70ºC
IP67
Non- Incendive
US/Canada:
Class I, II, III, Division 2, Group A, B, C, D, E, F, G, T4
Class I, Zone 2 AEx ia/nA [ia Ga] IIC T4 Ga/Gc
Class I, Zone 2 Ex ia/nA [ia Ga] IIC T4 Ga/Gc
Ta = -40ºC to +70ºC
Type 4X, IP67
ATEX – FM14ATEX0042X:
II 1/3 G Ex ia/nA [ia Ga] IIC T4 Ga/Gc
Ta = -15ºC to +70ºC
IP67
IEC – IECEx FMG 14.00018X:
Ex ia/nA [ia Ga] IIC T4 Ga/Gc
Ta = -15ºC to + 70ºC
IP67
Intrinsically Safe
US/Canada:
Class I, II, III, Div 1, Group A, B, C, D, E, F, G, T4,
Class I, Zone 0 AEx ia IIC T4 Ga
Class I, Zone 0 Ex ia IIC T4 Ga
Ta =-40ºC to + 70ºC
Type 4X, IP67
ATEX – FM14ATEX0041X:
II 1 G Ex ia IIC T4 Ga
Ta = -40ºC to +70ºC
IP67
IEC – IECEx FMG 14.0018X:
Ex ia IIC T4 Ga
Ta = -40ºC to +70ºC
IP67
Dust Ignition Proof
US/Canada:
Class II, III, Division 1, Group E, F and G, T4
Ta = -40ºC to +70ºC
Type 4X, IP67
ATEX – FM14ATEX0041X:
II 1/2 D Ex ia/tb [ia Da] IIIC T85ºC to T450ºC Da/Db
Ta = -15ºC to +70ºC
IP67
IEC – IECEx FMG 14.0018X:
Ex ia tb [ia Da] IIIC T75ºC to T435ºC Db
Ex ia IIIC T75ºC to T435ºC Da
Ta = -15ºC to +70ºC
IP67
The following approval standards are applicable:
FM3600:2011, FM3610:2010, FM3611:2004, FM3615:2006, FM3616:2011, FM3810:2005, ANSI/ISA60079-0:2013,
ANSI/ISA 60079-1:2009, ANSI/ISA 60079-11:2013, ANSI/ISA 60079-15:2012, ANSI/ISA 60079-26:2011, NEMA 250:2003,
ANSI/IEC 60529:2004, C22.2 No. 0.4:2009, C22.2 No. 0.5:2008, C22.2 No. 30:2007, C22.2 No. 94:2001, C22.2 No. 157:2012,
C22.2 No. 213:2012, C22.2 No. 1010.1:2009, CAN/CSA 60079-0:2011, CAN/CSA 60079-1:2011, CAN/CSA 60079-11:2011,
CAN/CSA 60079-15:2012, C22.2 No. 60529:2005, EN60079-0:2012, EN60079-1:2007, EN60079-11:2012, EN60079-15:2010,
EN60079-26:2007, EN60079-31:2009, A1:1991-2000, IEC60079-0:2011, IEC60079-1:2007, IEC60079-11:2011,
IEC60079-15:2010, IEC60079-26:2006, IEC60079-31:2008