Measurement accuracy
•
For the Samsung S7 tablet display, the battery is a 8000mAh Li-Ion type.
•
The AI Station HUB contains a Lithium Metal 3V / 45 mAh coin cell battery.
Measurement accuracy
Measurements reflect a physical property, such as area or volume for interpretation by a clinician. They do not
account for variations in acoustic speed of the body, which can vary significantly.
1
When used as per instructions scanning a bladder in a tissue equivalent phantom.
TABLE 6.
Measurement accuracy
Measurement
Units
Range
Accuracy
Modes
Distance
Axial (bladder)
cm
Full scale
±5% or ± 0.1cm
B-mode
Lateral (bladder)
cm
Full scale
±5% or ± 0.3cm
B-mode
Axial (vein)
cm
Full Scale
±5% or ± 0.1cm
B-mode
Lateral (vein)
cm
Full Scale
±5% or ± 0.1cm
B-mode
Volume
1
ml
0-999ml
±10% or ± 10ml
3D
Содержание EchoNous Bladder
Страница 1: ...P003948 007 Rev A March 2021 www EchoNous com User Manual ...
Страница 2: ...INTENTIONALLY LEFT BLANK ...
Страница 27: ...EchoNous Bladder Vein User Guide 21 Trademarks and document copyright EchoNous Inc 2020 ...
Страница 28: ...Important Information 22 EchoNous Bladder Vein User Guide INTENTIONALLY LEFT BLANK ...
Страница 64: ...Other Functions 58 EchoNous Bladder Vein User Guide INTENTIONALLY LEFT BLANK ...
Страница 78: ...Connectivity and Accessories 72 EchoNous Bladder Vein User Guide INTENTIONALLY LEFT BLANK ...
Страница 84: ...Specifications 78 EchoNous Bladder Vein User Guide INTENTIONALLY LEFT BLANK ...
Страница 101: ...EchoNous Bladder Vein User Guide 95 V version User Manual See User Manual W warnings 10 Z Zooming 44 47 ...