ECH (Changzhou) Medical Instrument Co., Ltd.
*
Use
10~40
°C
( 50~104
°F)
/RH(no-
condensing) / atmospheric pressure
range of 700~1060hPa
Measure position
Axillary (under the arm) – APP
display axillary temperature, which is
different from temperature
measurement taken orally.
Temperature Range
25.1~44.9
°C
(77.1~112.9
°F
)
Temperature Indication
Less Than 37.0
°C
(98.6
°F
) – gradient
from WHITE to GREEN
Equal to or greater than 37.0
°C
(98.6
°F
) - gradient from WHITE to
RED
Alert temperature (default alert
temperature in APP is 38.0
°C
(100.4
°F
) – WARNING LINE
Measurement Accuracy
±
0.1°C for 32-42°C, ±0.2°C for 25.1-
33.9°C/42.1-44.9°C(0.2°F for
89.6~107.6°F, ±0.4°F for 77.1-
93.1°F/107.7-112.9°F)
Display resolution
0.1
°C
(0.1
°F
)
Respond time
Once applied to the body, the device
need up to 8 minutes to reach a
stable reading
Transmission Distance
Up to 10 meters when applied to the
body
Water Ingress Protection
IP22 – Protected from damage due to
water dripping on the product
Standards
ISO 80601-2-56 standards for clinical
thermometer
IEC 60601-1 Medical Design
Standards, 3rd Edition
IEC 60601-1-2 Electromagnetic
compatible
Radio regulations
Bluetooth
®
Version 4.0
FCC Part 15
EN 300 328, EN 301 489-1/-17, EN
62479
Compatible Smart Devices
IOS 7.1 or later/ Android
TM
4.4 or
later based smart phones with
Bluetooth
®
4.0 interface. Visit
www.ech-med.com
for a list of
compatible devices.