ECH (Changzhou) Medical Instrument Co., Ltd.
temperatures, for example, when the
arm is raising exposing the sensor to
cold ambient temperature.
*
Confirm the patch thermometer is
correctly reading temperature by
measuring oral temperature with a
digital thermometer.
Temperatures from the patch
thermometer are higher than
expected
Confirm the thermometer is correctly
reading temperature by measuring
oral temperature with a digital
thermometer.
No audible alert is received from the
smart device when the patch is
detached or when a temperature alert
is generated.
*Check to insure that the phone
volume is set to an audible level and
is not set to mute;
*
Check settings for the Tempatch
TM
APP to insure that notifications are
enable and sounds is enabled for
notifications;
*
Note that the audible alert only
sounds for a short period. The visual
notification remains active in the APP
until cancelled by the user.
No notification is received when
temperature measured by the patch
thermometer exceeds the user set
alert level.
*Check to insure that the smart
device running the Tempatch
TM
APP
is in range of the patch thermometer.
No temperature alert will be
generated when the smart device is
out of range of the patch
thermometer.
*
Check to insure that the APP is
running. No alert will be generated if
the APP is not running.
11. Package List
Patch thermometer
(
Patch*1 for Child and Adult set / Patch*4 for Family Set
)
User manual (with warranty) *1
12. Warranty
The expired date of
Tempatch
TM
Wireless Intelligent Thermometer is two
years from production date.
Tempatch
TM
Warrants each new
Tempatch
TM
against defects in materials or workmanship within the expired date, and
agrees to replace any defective product without charge.
IMPORTANT: