amplitude should be less than 10mm
2.9
50Hz Anti-interference filter
≥
20dB
Table1 Input Impedance
Lead Electrode
The summit value of deflection front
traced by K open circuit (mm)
Lead Position
Connecting
to P1
Connecting to
P2
Single Channel ECG
Multi-channel ECG
Ⅰ
,
Ⅱ
,aVR, aVL,
aVF,V1
R
All other Lead
electrodes
Ⅰ
,
Ⅲ
,aVL, aVR,
aVF,V2
L
All other Lead
electrodes
Ⅱ
,
Ⅲ
,aVF, aVR,
aVL,V3
F
All other Lead
electrodes
Vi(i is 1
~
6)
Ci
All other Lead
electrodes
8
Vx, Vy, Vz
A,C,F,M
I, E, H
—
8
2.10
Features of Amplitude Frequency
When the filter is shut off, take 10Hz sine wave as reference.
From 0.5 to 50Hz, the tolerance of amplitude of frequency is -10%
~
+5%.
From 50 to 70Hz, the tolerance of amplitude of frequency is -30%
~
+5%
2.11
Features of Low Frequency .Time Constant no less than 3.2s
2.12
Baseline stability
2.12.1
Stable Power: baseline drifting should be no more than 1mm
2.12.2 Unstable Power: baseline drifting should be no more than 1mm
2.12.3 Sensitivity (no signal input) : baseline drifting should be no more than 2mm
2.12.4 Temperature Drift: From 5
~
40
℃
, baseline drifting should be no more than 0.5mm
/
℃
2.13 Paper Speed
Should have 25mm/s and 50mm/s at least , tolerance:
±
5%
2.14 Multi-channel Crosstalk for Multi-channel ECG
The Multi-channel crosstalk should less than 0.5mm between any channels of the multi-channel
ECG.
2.15 Influence of AC-DC converts
When the power supply is converted from AC to DC, the indicator light for DC work mode will on.
2.16 Printing resolution (Thermal matrix printing)