At the other function setup interface, use
【
▲
】
and
【
▼
】
button to select setting items among
“keypad tone”, “Beep volume”, “QRS Beep”, “Power-save” and “LCD backlight off time”, and use
and
button and number key to select and input. If user wants to keep the LCD backlight
always on or disable power-save function, set the value as 0.
Press
【
F1
】
button to save others setup and return, or press
【
F2
】
button to cancel and return.
Press F2 in the main interface to enter file management.
6 Cleaning, Disinfection and Maintenance
6.1 Cleaning
Turn off the ECG and remove the patient cable. Unplug the power cord from the AC outlet if
main supply applied.
1
)
Cleaning the Main Unit and Patient Cable
Wipe the exterior of the main unit and patient cable with a damp cloth using mild (non-caustic
neutral) detergent diluted in water. Then wipe it with a dry cloth.
2
)
Cleaning the Electrodes
Remove the residual gel from the electrodes with a clean soft cloth first. Take off the clip and
suction ball from electrodes connectors. Immerse them soap water (less than 35 Celsius) to remove
the gel, then use clean water to wash again, air dry or use a clean cloth to dry.
3
)
Cleaning the Printing Head
Dirty thermal printing head can deteriorate the record result. Clean the printing head at least one
time monthly.
Open the recorder cabinet and take out recording paper when the ECG is power off. Wipe the
printing head gently with a clean soft cloth damped in 75% alcohol. Do not use hard objects to