Ecel CURING LIGHT EC 500 Скачать руководство пользователя страница 4

Product Presentation

Packaging (single): Box (dimensions: 235mm x 190mm x 90mm)
List of items contained in the box: 01 Handpiece, 01 Charger Base, 01 Acrylic curing light
tip, 01 Eye protector, 01 Battery, 01 Charger and 01 Owner's manual.

1 - Light Guide Tip
2 - Eye Protector
3 - Activation button
4 - Handpiece
5 - Battery
6 - Led indicator of battery charge
7 - Charger Base
8 - Charger

1 - Photopolymerization resin.
Obs: The consumption materials have another ANVISA register and it is not supplied
with the equipment, it should be specified by the professional (Dentist).

Provide an electric outlet.
The power source is multi voltage, therefore it should be used in voltages from 100V to
240V with a frequency of 50/60Hz.
Unpack the equipment carefully.
Verify if all items that compose the equipment are present. (See section Accompanying
parts and accessories)
Connect the charger plug to the charger base.
Connect the charger (power plug) to the power supply.
Screw the battery on the inferior side of the handpiece.
Attach the acrylic curing light tip to the front orifice of the handpiece.
Plug the eye protector in the acrylic tip.
Set the handpiece on the charger base. Find a fresh and dry place, far from dust, humidity,
chemicals products and cleaning agents.

Charge the battery during 12 hours before the first use.

After attaching the tip to the equipment, it should be cleaned with some gauze, since
oiliness from the user's hands and other stains on the acrylic tip may affect light
conduction.

Install the equipment as described in the

section. Even on first use, the

equipment must be cleaned and disinfected, following the same procedures described in
the

section of this manual.

The Curing light equipment uses a blue light source for curing (photopolymerization)
resin.

is a wireless equipment, which allows freedom of movement

and total control. It is practical, portable, lightweight, and easy to clean and handling.
As an advantage, this equipment uses the latest LED technology, which replaces the

Accompanying parts and accessories:

Consumption material: (not sold with the equipment)

Curing Light EC 500

Installation

Cleaning and care

Curing Light EC 500

Installation

Product Description

!!

Procedure Before Product Use

4

Содержание CURING LIGHT EC 500

Страница 1: ...Owner s Manual CURING LIGHT EC 500 ...

Страница 2: ...IV O LT AG EM E 00 00 00 EC EL In dú st ria e Co m ér ci o Lt da Av M og ia na 18 54 Ri be irã o Pr et o SP Br as il CE P 14 07 5 26 0 PO TÊ NC IA 10 VA FR EQ 50 60 Hz NÚ ME RO DE SÉ RI E SE RI AL NU MB ER MO DE LO EC 50 0 TE NS ÃO MU LT IV OL TA GE M E 00 00 00 EC EL Ind ús tri a e Co mé rci o Ltd a Av Mo gia na 18 54 Ri be irã o Pr eto SP Br as il CE P 14 07 5 2 60 ...

Страница 3: ...090 848 0001 47 Mogiana Avenue 1854 Jd Independência Ribeirão Preto SP CEP 14075 260 Phone 16 3878 1700 Walter Bagatta Junior CREA SP 0601359652 Declared as exempt from registration by the Brazilian Health Department Figures Product Presentation Curing Light EC 500 Installation Procedure Before Product Use Product Description Operation Time Settings Charging the Battery Recharging the Battery Batt...

Страница 4: ...the acrylic curing light tip to the front orifice of the handpiece Plug the eye protector in the acrylic tip Set the handpiece on the charger base Find a fresh and dry place far from dust humidity chemicals products and cleaning agents Charge the battery during 12 hours before the first use After attaching the tip to the equipment it should be cleaned with some gauze since oiliness from the user s...

Страница 5: ... at this moment the LED will shut down keep pressing the button It will start a sequence of 10 tones with pauses of approximately one second between them Each tone corresponds to 10 seconds of the time setting Release the activation button after getting the desired number of tones which will correspond to the time working When the equipment is inserted in the charging base the battery indicator Le...

Страница 6: ...ld be checked the following items light intensity cracks and dryness in the power cable tension in the charger the tension of the battery charger time settings and general state of the light guide tip The supply of circuit diagrams list of parts components or other necessary information for Technical Assistance can be requested from the manufacturer and will be supplied by agreement This equipment...

Страница 7: ...quipment on patients with cardiac pacemakers Dispose the battery at an appropriate site according to local regulations avoiding environmental contamination and its inadequate use was designed to do not be sensitive to interferences as electromagnetic fields electric external influences electrostatic discharge or pressure and variations of pressure as long as the equipment be installed cleaned main...

Страница 8: ...525mmHg to 795mmHg ENVIRONMENTAL OPERATION CONDITIONS Atmospheric pressure range Environmental temperature functioning range 10 C to 35 C Environmental temperature range recommended by ECEL 21 C to 26 C Functioning relative humidity range non condensing 30 to 75 700hPa to 1060hPa 525mmHg to 795mmHg After Installation Environmental temperature range accommodation 5 C to 45 C Accommodation relative ...

Страница 9: ...TATION STORAGE ENVIRONMENTAL CONDITIONS Transportation and storage environmental temperature range 0 C to 55 C Transportation and storage relative humidity range non condensing 0 to 90 Atmospheric pressure range 500hPa to 1060hPa 375 mmHg to 795 mmHg The product must be transported and stored Carefully in order to prevent vibrations and impacts By keeping the side indicated by the arrow facing upw...

Страница 10: ...r Installation and or operation in disagreement with the instructions in this manual and falls Technician s travel expenses and equipment transportation expenses for technical assistance External finishing and or accessories damages not caused by the manufacturer Present the sales receipt to theTechnicalAssistance during the warranty Reasons for warranty voidance Items that are not covered by the ...

Страница 11: ...e battery again Solution Equipment does not charge Fault Cause Charger is disconnected or is not well connected Solution Verify if the charger is disconnected Equipment does not cure Fault Cause Obstruction of the light guide tip Solution Replace or clean it Cause There is no power supply Solution Await for power supply normalization or call for technical assistance Cause LED is burned out Solutio...

Страница 12: ...ECEL ABRIL 2009 Ecel Indústria e Comércio Ltda Fone 55 16 3878 1700 ecel ecel ind br www ecel ind br Av Mogiana 1854 Jd Independência Cep 14075 260 Ribeirão Preto SP Brasil ...