OPERATIONAL PROCEDURES 

A.  Preparing Standard Solutions 

Suitable conductivity standards are available commercially or the user can prepare them 
using research grade reagents. 

Here  are  some  standard  solutions  the  user  can  prepare  to  calibrate  the  probe  of  the 
model H10C. 

1.

 

Standard solution of 1413uS at  77°F (25°C): Accurately weight out 0.746 grams of 
research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water. 

2.

 

Standard solution of 12.90mS at 77°F (25°C): Accurately weight out 7.4365 grams of 
research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water. 

3.

 

Standard solution of 111.9mS at 77°F (25°C): Accurately weight out 74.264 grams of 
research grade dried Potassium Chloride (KCL). Dissolve in 1000ml of distilled water. 

[Note: You can store the remaining solution in a plastic container for one week but the 
air  space  between  the  cap  and  the  solution  must  be  kept  to  an  absolute  minimum. 
Storing  the  excess  solution  below  39°F  (4°C)  can  increase  the  storage  life.  If  you  have 
any doubt of the accuracy of the stored solution, a fresh batch should be prepared.

B.  Calibration 

Calibration  setup  contains  five  sections:  TDS,  Cell,  Temperature  Coefficient, 
Temperature reference and Conductivity Calibration. To access these sections: 

1.

 

Connect the conductivity probe, PN 2008210, to the unit and turn the unit on. The 
screen will display CELL and the cell constant of the conductivity probe. 

2.

 

Allow  temperature  reading  to  stabilize,  press  “Cal”  key  to  enter  the  calibration 
mode.  CAL  appears  on  the  LCD.  Press  “Mode”  key  to  sequentially  display  the 
following sections: 

[Note: Press Enter key to accept any values changes in each section and automatically 
advance to the next section. If there are no changes, the unit accepts the current value 
and proceeds to the next section.

a.

 

TDS 

TDS is determined by multiplying conductivity (mS) by a TDS factor. The default factor 
value is 0.65. To change the TDS factor, use the up and down keys to adjust the value 
between 0.30 and 1.00. Press “Enter” key to save the new value, or press “Mode” key to 
cancel the change and display the CELL screen. 

Содержание H10C

Страница 1: ...Instruction Manual Microprocessor based Conductivity TDS Salinity Temperature Hand Held Meter Electro Chemical Devices Inc 1500 N Kellogg Dr Anaheim CA 92807 USA Telephone 1 714 695 0051 www ecdanaly...

Страница 2: ...ays and Key Functions 4 Keys 5 OPERATIONAL PROCEDURES 6 A Preparing Standard Solutions 6 B Calibration 6 a TDS 6 b CELL 7 c Temperature Coefficient 7 d Temperature Reference 7 e Conductivity Calibrati...

Страница 3: ...ts features include automatic conductivity ranging automatic temperature compensation long battery life and 50 60 Hz AC noise rejection This unit is universal in application and user friendly for fiel...

Страница 4: ...ttery cover to expose the battery compartment Figure 1 2 Note the polarity and insert the batteries into the battery compartment correctly 3 Replace the battery cover and make sure to secure the two s...

Страница 5: ...uishes between temperature compensated and uncompensated readings 10 ppt Parts per thousand indicates salinity measurement 4 MAIN DISPLAY Displays conductivity salinity and TDS numerical values 11 uS...

Страница 6: ...dissolved solids TDS and uncompensated conductivity In calibration mode this key exits the current calibration and displays the next calibration parameter Up Down Increases or decreases the displayed...

Страница 7: ...solution below 39 F 4 C can increase the storage life If you have any doubt of the accuracy of the stored solution a fresh batch should be prepared B Calibration Calibration setup contains five secti...

Страница 8: ...mperature Reference screen d Temperature Reference The unit uses the temperature reference value to calculate temperature compensated conductivity The default value is 77 F 25 C To change the temperat...

Страница 9: ...Cause Corrective Action over during Measurement Conductivity is 200mS Salinity is 70 0 ppt Ensure the probe is immersed at least 2 3 or 5 7cm from the tip below the surface of the sample standard sol...

Страница 10: ...14 F to 194 F 10 0 to 90 0 C 0 1 C 0 2 C or 0 4 Full Scale whichever is greater Reference Temperature 59 to 77 F 15 0 to 25 0 C Temperature Coefficient 0 0 to 4 0 Cell Constant 0 5 0 15 TDS Constant R...

Страница 11: ...ment shall be limited to a refund in the purchase price or at ECD s discretion to the repair or replacement of any product that proves upon ECD s examination to be defective when returned to the facto...

Страница 12: ...1500 N Kellogg Dr Anaheim CA 92807 Phone 1 714 695 0051 www ecdanalytical com...

Отзывы: