eccola H-MAN X 2 Скачать руководство пользователя страница 4

 

Operating instruction  04_2022 

Page 4 

Service Address: 

eccola GmbH

  -  Im Alber 8 -  D-73084 Salach  -  Phone.:0049-7162-948996 

Internet: 

www.eccola.de

  email: [email protected] 

 

Different hearing test procedures 
 

The different possibilities of the hearing test procedure 
 

There are 2 different possibilites to run the hearing test. 

 

I. 

The hearing test runs automatically until the end 

Setting prerequisites: DIP switch 4 on “off” and DIP switch 1 on “off” 
 
The hearing test is started by the test person pressing the start key. After this single 
pressing of the start key the test runs automatically until the end – i.e. 500 Hz, 1000 Hz, 
2000 Hz and 4000 Hz right and left. At the same the test person can enter the result in the 
reply card. This means that the hearing test can be made without any assistant or 
supervisor and is suitable for tests in front of the shop, in shopping malls, pharmacies and 
during promotion days. The result is self-explanatory by means of the result card. 

 

II. 

Interactive hearing test with confirmation by pressing the start 
key  

Setting prerequisites: DIP switch 4 on “on” HOLD-LED function with “cycle continuing” 
function.

 

 
The hearing test is started by pressing the start key for the first time. As soon as a tone 
will be heard and the corresponding LED display will flash, the repeated pressing of the 
start key will allow to confirm that the test person has heard the tone. The LED of the 
heard tone remains flashing, showing that the tone has been heard. Then follows the next 
tone and the procedure is the same as described for the first tone. 
After the last tone the start key will be without function during 4 seconds thus showing the 
result by means of the flashing LEDs (tones heard). 
Thereafter the result is shown during 25 seconds provided that the start key is not 
pressed. This means that the “heard” result can be entered in the result card and/or 
discussed wih the test person.  The result will disappear after 25 seconds. 
As soon as the start key is pressed again after 4 seconds, the current result will be 
deleted and the test can start again from the very beginning. 
 
This adjustment if particularly suitable for hearing test campaigns with promoters or 
assistants. This facilitates initiating a conversation with  a potential new customer.  
This test can of course also be performed by the test person on its own by means of the 
attached manual. 

 

 

Содержание H-MAN X 2

Страница 1: ...r location you may readjust the sound level see functional characteristics Functional Characteristics Important to know The hearing tester is operated by means of batteries 4 pcs LR6 of 1 5 V each which allows its use in any location The test frequencies are 4 tones on either side 0 5 1 2 4 kHz You only need to remove the back cover panel in order to customize the H MAN X 2 For this purpose gently...

Страница 2: ...rnal DIP Switches Master Master 1 5 Volume level Volume level 1 2 3 4 5 0 5 1 2 4 Sound level kHz kHz kHz kHz Frequencies 1 2 3 4 0 5 1 2 4 kHz Headphone jack LEDs Power supply jack Battery Start Counter 9 12V 4x1 5V Button In case of additionally using a counter you will have the possibility to reset the same to 0 by means of a magnet included in the delivery For this purpose you only need to mov...

Страница 3: ... of the hearing tester can be continued without any problems In the event of a battery voltage of less than 5 V it is no longer possible to start the hearing tester Power input during operation about 11 mA After the automatic stop of the hearing tester the power input is less than 1µA multimeters normally in use do not allow to realize any measurements We recommend alkaline manganese batteries wit...

Страница 4: ... LED function with cycle continuing function The hearing test is started by pressing the start key for the first time As soon as a tone will be heard and the corresponding LED display will flash the repeated pressing of the start key will allow to confirm that the test person has heard the tone The LED of the heard tone remains flashing showing that the tone has been heard Then follows the next to...

Отзывы: