For assembling put the ear wax filter into the housing. put on cover and
close it clockwise by 30 degrees (slide unlocker to the right).
The part
with the moveable cover must again be on the right side when installing
the cover. The filter is placed correctly when the arrow on the filter
housing is pointing in the same direction(to the left) as the arrow on the
housing.
Then attach the transparent hoses
on the right and left side.Make sure
that the filter is tightened correctly
when inserting it. This can be
checked by switching on the device
and closing the red hose with your
finger. The LED should then change from green to red (vacuum). In case
the LED remains green the filter has either not been placed correctly,
the cover has been put on incorrectly or the hoses have not been
inserted correctly. Please repeat the whole process. From time to time it
may happen that the cleaning attachments are dirty. This can be seen
when the red LED (vacuum) lights up permanently instead oft he green
one. The attachments can also be cleaned by means of a needle – or we
recommend to clean the attachments in an ultrasonic bath and to check if
the hoses are correctly tightend on the back oft he device.In case
sucking function is still not working you should check the filter. It might be
dirty or filled with ear wax. Then please clean the filter.
Please clean the hose with the metal piece on the hose tip regularly
(once a week) – you can remove it and clean it in the ultrasonic bath,
leave it to dry overnight and reattach it.
The metal handpiece can also be cleaned separately by detaching the
metal handpiece on the hose tip by a rotary movement, then clean it and
reattach it on the hose by a rotary movement.