Under normal working conditions, your Reverse Osmosis (RO) water filter will not 
have any microbiological matter (bacteria) growing inside it. Unexpected extraneous 
conditions (eg temporary local water treatment problems, static water due to lack of 
use, poor maintenance etc) can result in bacteria growing inside your RO unit. Left 
unchecked, this bacteria can be detrimental to health in a clinical application and 
needs removing.
We recommend MONTHLY testing of the water dispensed from your RO unit with a 
TVC (total viable count) 

dipslide

 to check whether bacteria has grown inside the RO 

unit. In the unlikely event of your microbiological test detecting bacteria, we supply 
a simple ‘

RO cleaning kit

’ (supplier code EY90200-ST), which contains components 

needed to clean the RO system. In addition, you will need a 5L bottle of ‘

Cl

’ 

biofilm cleaner, all available from your local supplier.
A video for how to run the procedure is available online on YouTube, simply type in 
‘Clean Eau Flow 505 RO’ in the search bar.

Biofilm ‘cleaning kit’

UK HTM01/05- sec 6.50:

 ‘Specialist items of equipment...may require cleaning and 

decontamination processes that are purpose designed’.

Содержание 505

Страница 1: ...05 Hard Water code EY5505HW Installation Manual Thank you very much for selecting an Eau Flow In order to bring the out best use of your system please read this instruction manual carefully before ins...

Страница 2: ...return valve EY20110 B Appliance tap EY20111 C to BSP x3 EY20112 D 6 2mm tube 10m EY20008 E RO unit EY5505HW F Tank ball valve EY20127 G Storage tank 12L EYST12 H T Piece Push fit EY20123 I Dispensing...

Страница 3: ...2 has any plumbing deadlegs in the plumbing between the stop cock in the road and the RO unit plumbing point Failure to observe these conditions will result in bacterial contamination of your RO syst...

Страница 4: ...ges and run from dispensing point valve or tap for 5 minutes down the drain 5 Now close the dispensing point valve or tap and open the valve on top of storage tank inline with tubing 6 Allow storage t...

Страница 5: ...intenance The following periodic maintenance is recommended so your system will provide an uninterrupted supply of purified water HARD WATER areas 250ppm CaCO3 please ensure you add a post RO blue dei...

Страница 6: ...than 9 months since you changed cartridges then you need an additional blue Di cartridge EY30810 plus inline TDS meter EY20155 between RO and dispensing valve tap Contact your rep for details Power l...

Страница 7: ...ion and needs removing We recommend MONTHLY testing of the water dispensed from your RO unit with a TVC total viable count dipslide to check whether bacteria has grown inside the RO unit In the unlike...

Отзывы: