XS series operating manual
XS user manual v0.55
- 87 -
Multiple ground continuity followed by a dielectric strength test ( automatic
sequence)
Fig 10.15
The case of a multiple continuity immediately followed by a strength test requires a precise care.
The multiple ground continuity function means that the user sometimes enters the security area or is
in direct contact with the specimen under test.
For that reason, in order to protect the user's safety, the unit stops the proceeding of the test before
starting the strength test.
At this stage, test can be released by :
1. T
he keyboard :
Press [ENTER] and then [LOW ARROW] within 0.5 sec. and keep this last key pressed until it is
taken into account.
2.
Using the safety loop :
If the safety loop is open during or at the end of the multiple continuity test, closing of the safety
loop will start the strength test.