4.
USING T CARE COMPACT
For the correct use of the device it is necessary to:
1. Select the capacitive (CAP) or resistive (RES) operational mode
2. Programming the functional parameters
3. Connect the electrodes to be applied on the area to be treated.
The operating parameters can be set manually from the operator interface (
touch-screen
).
Recommended treatment parameters for some of the most common pathologies can be
found in the following Table:
PATOLOGY
DURATION
MODALITY
FREQUENCY
POWER
LOW BACK PAIN
20 MIN
CAP 50%- RES
50%
500 KHz
CAP 30-40%
RES 25-35%
LUMBAR
HERNIATED DISC
20 MIN
CAP 50%- RES
50%
500 KHz
CAP 20-30%
RES 10-15%
EPICONDYLITIS
20 MIN
CAP 50%- RES
50%
500 KHz
CAP 20-25%
RES 15-20%
GONARTROSIS
20 MIN
CAP 50%- RES
50%
500 KHz
CAP 20-30%
RES 15-25%
ACUTE
ACHILLES
TENDINOPATHY
15-20 MIN
CAP 100%
500 KHz
5-10%
CHRONIC
ACHILLES
TENDINOPATHY
15-20 MIN
CAP 50%- RES
50%
500 KHz
CAP 25-35%
RES 20-35%
Energy delivery can be made in two way:
1) Continuous
2) Pulsed (500 kHz)
4.1
Knob for power adjust
Using the knob
on the front panel
, you can increase
or decrease
the value of the
power to
be delivered
(in % of maximum value); the changed value is shown on the display.
On the knob there is also a key for
start / pause
command.
4.2
Operations with user interface
When the device is switched-on, for about 2 seconds the logo
Easytech
and indication
T
CaRe Compact
are displayed.