background image

 

Easytech s.r.l. –T CaRe Compact User’s Manual 

     

rev. D 

Pag. 1 di 36

 

 

 

T CaRe Compact 

Mod. 80000650 

 

User’s Manual 

 

 

 

                                                                   

EASYTECH s.r.l. 

Via della Fangosa, 32 – 50032  Borgo San Lorenzo  (FI) – Italy 

tel. 055 8455216 – fax 055 8454349 

e-mail: 

[email protected]

 

Direct line of  

Easytech technical support

+39 348 2316323

 

 

      

    

             1936

 

Содержание T CaRe Compact

Страница 1: ...Pag 1 di 36 T CaRe Compact Mod 80000650 User s Manual EASYTECH s r l Via della Fangosa 32 50032 Borgo San Lorenzo FI Italy tel 055 8455216 fax 055 8454349 e mail info easytechitalia com Direct line o...

Страница 2: ...tten authorization from Easytech s r l via della Fangosa 32 Borgo San Lorenzo Florence tel 39 0558455216 fax 39 0558454349 Easytech s r l reserves the right to revise this publication and make changes...

Страница 3: ...ical Characteristics 20 4 USING T CARE COMPACT 21 4 1 Knob for power adjust 21 4 2 Operations with user interface 21 4 3 Treatment 22 4 4 Manual Settings 22 4 5 Set up commands 25 5 GUIDELINES FOR TRE...

Страница 4: ...the condenser among which the biological tissue is interposed becoming an electrical conductor of the second class The energy supply provided by the generator produces a concentration of electrical ch...

Страница 5: ...dical devices Part 1 Evaluation and testing within a risk management process EN ISO 15223 1 Medical Devices Symbols to be used with medical devices labels labelling and information to be supplied Part...

Страница 6: ...y complies with that indicated on the device label Before installing the device verify that the packaging and its content are not damaged Only use the power cord supplied with T CaRe Compact or the sp...

Страница 7: ...and in general relevant rehabilitative treatments in order to understand the potential of the device in question An adequate knowledge of the local language is required Only slight visual or sensory...

Страница 8: ...60601 1 2 and comprehensive of all types of electromagnetic interference that may cause incompatibility between several units in a given location The first table concerns the emissions of T CaRe Compa...

Страница 9: ...nded for use in the electromagnetic environment specified below The customer or the user should assure that it is used in such an environment Immunity test Test level EN 60601 1 2 Compliance Level Ele...

Страница 10: ...ixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked wit...

Страница 11: ...nd reflection from structures objects and people a Field strength from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM ra...

Страница 12: ...t power P of transmitter W From 150kHz to 80MHz d 1 2 P From 80MHz to 800MHz d 1 2 P From 800MHz to 2GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For tr...

Страница 13: ...ine in the case of devices for applications other than cardiology For patients with hypotension long term applications or applications on the cervical vertebrae or spine may cause low arterial pressur...

Страница 14: ...ust supervise the treatment during various stages in order to guarantee its prompt intervention in case of need the operator must use the equipment in accordance with the instructions contained in thi...

Страница 15: ...ions It is equipped with A touch screen This is the operator interface for controlling the device which allows the selection of the operating mode the operating parameters treatment activation B a pow...

Страница 16: ...surface EasyTech provides electrodes of different shape and size Resistive electrodes a circular electrodes with a flat surface and three different sizes of diameter 35mm 55mm 75mm b cylindrical wedge...

Страница 17: ...6 Figure 3 2 Capactive Electrode Figure 3 3 Resistive Electrode Figure 3 4 Return Plate Passive Electrode Figure 3 5 Elbow pipes The optional electrodes are shown in the figures 3 6 3 7 3 8 Figures 3...

Страница 18: ...ode and the skin and between skin and return plate it s valid only for the steel return plate because the disposables ones are pre soaked with a conductive gel Easytech invites to use the type of crea...

Страница 19: ...supplied n 1 T Care Compact Unit n 1 User Manual n 1 Resistive Handpiece n 3 Resistives Electrodes complete with cover D 75 D 55 D 35 n 1 Capacitive Handpiece n 3 Capacitive Electrodes complete with c...

Страница 20: ...ghtness 130x72mm Working frequencies 500 kHz 5 Output power RES at 82 ohm 200 W Output power CAP at 1000 ohm 250 VA Maximum output voltage on the resistive electrode open circuit 340 Vpp Maximum outpu...

Страница 21: ...35 LUMBAR HERNIATED DISC 20 MIN CAP 50 RES 50 500 KHz CAP 20 30 RES 10 15 EPICONDYLITIS 20 MIN CAP 50 RES 50 500 KHz CAP 20 25 RES 15 20 GONARTROSIS 20 MIN CAP 50 RES 50 500 KHz CAP 20 30 RES 15 25 A...

Страница 22: ...ou select see Figure 4 1 and Figure 4 2 resistive or capacitive mode of operations the energy output mode You can have continuous mode symbol or pulsed mode symbol the working frequency 500 kHz the tr...

Страница 23: ...area to be treated 2 give the start command 3 Set the power to be delivered1 in of the maximum The start command can also be given through the button on the knob When the treatment begins see Figure 4...

Страница 24: ...nnot tolerate the heat generated with the set power level eg for simple hypersensitivity you can put the unit in Pause press in order to stop the timer and change the setting Reached the new value pre...

Страница 25: ...power language setting Italian English French German Spanish Dutch Portuguese other volume setup for sounds produced during treatment The adjustment can be made on 8 levels display contrast adjustmen...

Страница 26: ...3 How to place the Electrodes The position of the reference electrode and the mobile one depend on pathologies to be treated a layer of conductive cream must always be used between the surface of eac...

Страница 27: ...e effect of resistive diathermy is to generate heat in deep areas therefore it is impossible to measure the temperature achieved for this reason we must consider the patient s response The mobile elec...

Страница 28: ...flat on the patient s skin to avoid a concentration of current on a small area which could have harmful effects Treatment is started with a low power level which is then increased until the patient do...

Страница 29: ...ant damage before it takes place the normal patient s defensive reaction distancing from the applicator In case of failure the only instance in which the risk may be modest but important for the patie...

Страница 30: ...c rule it is recommended an annual inspection by a specialized laboratory which can also make the performance at home The interval of a year is the most commonly used for devices of this class however...

Страница 31: ...etting consists instead in a series of actions to ensure that the machine measure correctly again when the error is not caused by the presence of a fault but only by a drift or by the replacement of a...

Страница 32: ...C Relative Humidity 0 75 Atmospheric Pressure 700 1060 mbar 8 2 Packaging In the case of shipping by courier you should create a package that can guarantee protection from shocks or falls There are n...

Страница 33: ...itch Non ionizing radiation It is located on the device plate Type BF Applied Part Classification According CEI EN 60601 1 Code This number is just an example Read User Manual The device that is no lo...

Страница 34: ...Easytech s r l User Manual T CaRe Compact rev D Pag 34 di 36...

Страница 35: ...Easytech s r l User Manual T CaRe Compact rev D Pag 35 di 36...

Страница 36: ...r l Via della Fangosa 32 50032 Borgo S Lorenzo Firenze Italy Tel 39 055 8455216 Fax 39 055 8454349 E Mail info easytechitalia com www easytechitalia com Direct line of Easytech technical support 39 3...

Отзывы: