Easytech s.r.l. – User Manual T CaRe Compact
rev. D
Pag. 8 di 36
2.4
Warnings regarding electromagnetic interference
T CaRe
Compact
produces electromagnetic energy for emitting it for therapeutic purposes.
The device is manufactured in accordance with the laws and the standards in force and, if
used according to guidelines included in this manual, it radiates only very small quantities of
electromagnetic energy like any other electronic device not used for telecommunications.
Normally, therefore, it will not disturb other equipment nor produce radiation potentially
dangerous for people, whether they are the patient, the operator or third parties.
T CaRe
Compact
is manufactured so as to be immune from noise due to other equipment,
provided that the latter complies with the standards and, in the case intentionally it emits
electromagnetic radiation (for therapeutic use or for telecommunications), it is maintained a
suitable distance to avoid interference.
2.5
EMC tables
What mentioned above in a qualitative way, is better specified and quantified by the
following tables, prescribed by the standard EN 60601-1-2 and comprehensive of all types of
electromagnetic interference that may cause incompatibility between several units in a given
location.
The first table concerns the emissions of
T CaRe
Compact
to other devices; the other three
relate to the ability of
T CaRe
Compact
not to be affected by external disturbances; the last
table shows the separation distance required between RF communication devices and
T
CaRe Compact
. On the contrary a significant degradation in device performance may occur.
The following Warning should be noted:
"Portable RF communication devices (including peripheral devices such as antenna cables
and external antennas) should not be used within 30 cm (12 inches) from T
Care Compact
,
including cables. Otherwise, the device may suffer degraded performance. "
The following tables are based on the test results, according to clause 6 of the EN 60601-1-2
standard.
In relation to the essential performances, the device does not provide immediate support for
the patient, certainly not life support. Therefore, there are no essential "therapeutic"
performances to be maintained in critical situations.
WARNING
T Care Compact is a Class A product. In a residential environment this device may cause
radio interference in which case the user may be required to take adequate
countermeasures.