INFLATE TIME
REST TIME
touch
6
8
7
5
17
Alarms and Alerts
The Toto Touch control unit has three integrated alarms and one information alert to ensure patient
safety and efficacy.
The sound pressure level of the alarm audio or the reminder audio signal ranges from 55dB-85dB.
The alarm frequency is 150HZ to 1000HZ.
While troubleshooting, the alarm can be muted for up to 10 minutes. If the fault is not corrected
during this time, the alarm reactivates.
If a new alarm condition occurs within 10 minutes, mute will cease, and the new condition will
generate an alarm.
Please Note:
•
The corresponding alarm indication LED remains lit when the alarm is muted.
•
If the alarm condition is corrected while muted, the LED and mute will automatically terminate.
•
The alarm audio signal comes from a speaker, and the remainder audio signal comes from
a buzzer.
•
Alarm activation may take up to two minutes from when the alarm state occurs.
Alarms and Alerts
Alarm Condition
Indicated by an audible alarm and:
Air Leak Alarm
The pressure in the system
has fallen below the minimum
operational requirements, i.e.
an air leak.
Low priority
Air Leak LED permanently lit yellow
(when activated) (6).
Blockage Alarm
Airflow between the control unit
and the platform is prevented, i.e. a
blocked air hose.
Low priority
Blockage LED permanently lit yellow
(when activated) (7).
Service Interval
Service required; default at one-year
use.
n/a
Service LED permanently lit amber
(8) (NO AUDIBLE alarm).
Power Loss Alarm
Power to the control unit has been
interrupted.
Low priority
Audible alarm, no LED lit.
Initialisation Failure
Control unit fails to start
n/a
All ‘Inflate time LEDs’ and ‘Rest time
LEDs’ flashing (when activated).