ADJUSTMENTS AND REPLACEMENTS
Additional Service Procedures
06DEC05
AR4366-1
Page
202 of 219
Setting the Laser Calibration Voltage
Important
•
You must do this procedure when installing a new OPTICAL AY.
•
For systems with software version
≥
4.1,
you must have a “Session ID” for access to
“Service Functions” and “Diagnostics”. See SERVICE BULLETIN 843.
1
At the main menu, click:
•
[Service Functions]
•
[Diagnostics]
•
[LASER POWER]
2
In the “Laser Calibration” window, enter the values for the new F-THETA LENS.
3
Click:
•
[CALIBRATE NEW LASER]
•
[SAVE NEW BACK UP]
4
Doing a Calibration for the Scan
for all size PLATES.