Eastern Energy DT-619 Скачать руководство пользователя страница 4

Operation 

Connecting the Vane 

1.  The vane plug is inserted in the meter’s sensor jack at the top of the meter. The plug 

and jack are keyed so that the plug can only fit in the jack one way. 

2.  Turn the plug carefully until it lines up with the jack and then firmly push the plug in 

place. Do not apply undue force or try to twist the plug side-to-side. 

3.  If the vane is not connected to the meter or if the sensor is defective, the LCD display 

will indicate 

OL

 in place of a Temperature reading. 

 

Side view of Vane 

Air Velocity Measurements (Single Point) 

1.  Turn on the meter using the ON/OFF      button. 
2. Press 

UNITS

 button to select the desired unit of 

measure. 

NOTE:

 At power up the meter will 

display the last unit of measure previously 
entered. 

Arrow 

 

3.  Place the sensor in the air stream. Ensure that 

the air enters the vane as indicated by the arrow 
sticker placed inside the vane. 

4.  View the readings on the LCD Display. The 

large main LCD display shows the Air Velocity 
reading. The upper right LCD sub-display shows 
the temperature reading. 

Air Velocity Averaging Mode 

1.  To enter 20 Point Averaging Mode, press and hold the 

AVG

 button until it beeps twice. 

The 

AVG

 icon will be displayed. 

2.  Take a measurement and press the 

AVG

 button. A single beep will sound and the 

HOLD

 icon will appear in the display. 

3.  The average reading will be displayed and number of readings measured will appear in 

the upper right hand corner of the display. After 5 seconds, the display will return to the 
current reading. 

4.  Repeat steps 2 - 3 until the desired number of points have been measured. 
5.  To return to standard velocity measuring mode press and hold 

AVG

 button until meter 

beeps twice. 

Note:

 In the standard velocity measuring mode, press the 

AVG

 button once to recall the 

previous average. The average will be cleared when you enter the Averaging Mode 
again. 

airflow 

Eastern Energy Co., Ltd.

 

40/4 Vitoondumri Rd., Banbueng

   

Banbueng, Chonburi, Thailand

 

 

Tel: 66-3844-6117

[email protected]

www.eastern-energy.com

www.eastern-energy.net

Содержание DT-619

Страница 1: ...User s Manual CFM CMM Thermo Anemometer Model DT 619 Eastern Energy Co Ltd 40 4 Vitoondumri Rd Banbueng Banbueng Chonburi Thailand Tel 66 3844 6117 sale ete co th www eastern energy com...

Страница 2: ...e number button in AREA mode AVG Used to Average multiple readings in FLOW or VELOCITY mode Up to 20 readings can be averaged HOLD Press to freeze the displayed reading Press again to unlock display T...

Страница 3: ...rature units of measure CFM CMM airflow units of measure Ft 2 m 2 units for area dimensions m s ft min km h MPH knots air velocity units of measure X10 X100 multipliers for air flow readings AVG air a...

Страница 4: ...he LCD Display The large main LCD display shows the Air Velocity reading The upper right LCD sub display shows the temperature reading Air Velocity Averaging Mode 1 To enter 20 Point Averaging Mode pr...

Страница 5: ...the desired location Once you have selected the desired memory location enter your dimension Use the UP button to change the flashing digit Use the LEFT button to move the decimal Use RIGHT button to...

Страница 6: ...rrent readings NOTE the meter will keep recording MAX MIN AVG readings 5 To clear and stop MAX MIN AVG recording and return to normal operation press and hold the MAX MIN button until the meter beeps...

Страница 7: ...g per second approx Sensors Air velocity flow sensor Conventional angled vane arms with low friction ball bearing Temperature sensor NTC type precision thermistor Automatic Power off Auto shut off aft...

Страница 8: ...x Area ft2 NOTE Measurements made in inches must be converted to feet or meters before using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 2...

Отзывы: