14
6 Advanced Application
6.1 Setting 0 % and 100 % output
parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. Some level values have been set when it is out of the factory and set values are
shown as below:
Output function
0% value
100% value
Resistance 400Ω
0.00 Ω
400.00 Ω
Resistance 3200Ω
0.0 Ω
3200.0 Ω
Pt100
0.0
℃
500.0
℃
Pt1000
0.0
℃
400.0
℃
Cu50
0.0
℃
150.0
℃
Cu100
0.0
℃
150.0
℃
The set values may not meet your requirements, so which can be reset. Then you can
display the percentage with stepping or slope output as well as display the percentage.
Adjust the output value with keys
, press
or
for a long time until the buzzer
works to enter into 0% or 100% values. The new defined range value is saved in the
calibrator storage automatically and remains effective after restart.
The flowing
operations are available with the setting:
Press
or
for manual steeping (increase or decrease) output based
on 25% increment.
Press
or
instantly to make output jump within the range of 0% to
100%.