![East Tester ET-1712 RTD Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/east-tester/et-1712-rtd/et-1712-rtd_user-manual_522218009.webp)
6
3.
Then the type of product and automatic shutdown time will be displayed for 2
seconds as shown in Figure 2.1-3.
Figure 2.1-3
4.
Press
and
to switch to the thermal resistance output mode. The
screen will display as shown in Figure 2.1-4.
Figure 2.1-4
5.
Press
and
to increase or decrease 1 of the horizontal line position (the
number will be based on automatic stepping without change of horizontal line
position); press
or
to change the position of horizontal line.
6.
Press
until the buzzer works to enter 0.0
℃
as the 0% value.
7.
Likewise, press
to increase output value to be 100.0
℃
, and then
press
until the buzzer works to enter 100.0
℃
as 100% value.
8.
Press
or
to make output increase or decrease within the range of
0% to 100% based on 25% advance amplitude. The screen will display as shown
in Figure 2.1-5.