background image

 

The material contained in this document is the property of Electronics & Innovation Ltd., it is 
subject to change without notice.  
 
August 2012 

Revision E 

 

 

Table 1-1. SPECIFICATIONS 

 

FREQUENCY COVERAGE: 

300 kHz to 35 MHz. 

GAIN: 

55 dB min, ±1.5 variation. 

CLASS A LINEAR OUTPUT: 

150 watts. 

HARMONIC DISTORTION: 

< -25 dBc at 125 watts output. 

SATURATED RF POWER OUTPUT: 

250W Min. 

from  300kHz - 4MHz; 

200W Min. 

from  4 MHz - 35MHz; 

INPUT IMPEDANCE 

50 ohms, VSWR, 1.5:1 Maximum. 

OUTPUT IMPEDANCE: 

50 ohms, VSWR, 2:1 Maximum 

STs  STABILITY: 

Continuous operation into any load or 
source impedance. 

PROTECTION: 

Unit will withstand a + 13dBm input 
signal (1.0 Volts RMS) for all output load 
conditions, without damage. 

POWER OUTPUT METER: 

True RMS power detection.  ±3% of full 
scale accuracy. 

POWER REQUIREMENTS: 

100 

– 240 VAC 47-63 Hz. 

SIZE: 

5.3

” x 16.5” x 18.4” 

134.5 x 420 x 467 mm 

WEIGHT: 

52.5 pounds 23.8 kg 

CONNECTORS: 

BNC 

OPERATING   TEMPERATURE: 

– 40 C 

RACK  MOUNTING: 

 

 
 
 

Chapter 2 Operation 

 
2.1 INTRODUCTION 

 
The  A150  RF  amplifier  is  used  to  amplify  the  RF  level  of  signal  sources  in  the 
300 KHz to 35 MHz range. No tuning or any other form of adjustment is required.  
 
The A150 produces rated power output at its output connector, regardless of load 
impedance. Any power reflected due to output load mismatch is absorbed in the 
amplifier.  Therefore,  although  the  output  impedance  is  50  ohms  (maximum 
VSWR: 2:1), the amplifier will work into any load impedance. 
 
 

Содержание A150

Страница 1: ...hout notice August 2012 Revision E 1 A150 Broadband Power Amplifier HIGH RF VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these volt...

Страница 2: ...lly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cos...

Страница 3: ...M FM SSB pulse and any complex modulation signals The amplifier has 55 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is...

Страница 4: ...m input signal 1 0 Volts RMS for all output load conditions without damage POWER OUTPUT METER True RMS power detection 3 of full scale accuracy POWER REQUIREMENTS 100 240 VAC 47 63 Hz SIZE 5 3 x 16 5...

Страница 5: ...1 V rms indicated maximum input voltage is 5 times the level of the input signal required to achieve maximum output Input voltages in excess of 2 volts peak may permanently damage the instrument iii...

Страница 6: ...icted If the fault persists please contact Field Service 2 3 3 RS 232 Interface The A150 features a standard RS 232 serial interface suitable for connection to a PC or host system The communication pr...

Страница 7: ...nt end The output signal of the MMIC is coupled to the gate of transistor Q1 The further amplified signal appearing at the drain of Q1 is coupled to the input of Q2 This is transformed to 50 and fed t...

Страница 8: ...AC_line 110VAC_neutral earth_gnd SELV 5V Inhibit SELV 5V ret Inhibit_ret DC_ok DC_ok_ret 48V_ret 24V 24V_ret 48V A1 Fan 24V Ret 1 12 05 P1 5 9 4 8 3 7 2 6 1 J2 24V_ret 48V_ret RS232 Control Port J5 AC...

Страница 9: ...a power level of 13dBm This chapter therefore deals only with certain fundamental procedures for fault location S1 VHF_splitter_20W RF_in RF_out1 RF_out2 50 ohm 34dBm 48V_ret 48V_ret 24V_ret 2100L SC...

Страница 10: ...or c Adjust the output level of the sweep generator so that a 50 ohm video detector connected at the output of the unit will not be damaged by excessive power output Reference section 4 4 1 for set up...

Страница 11: ...in Measurement 2 Connect the equipment as shown in Figure 4 1 then proceed as follows a Set the oscilloscope to DC Time cm to Ext X and gain to l0mv cm b Set the sweep generator to the video sweep mod...

Страница 12: ...rsus frequency sweep on the oscilloscope The average gain should be greater 50 dB The gain variation should be within the 3 dB as shown on the oscilloscope 4 4 2 Measurement of Harmonics Figure 4 2 1...

Страница 13: ...prevent movement Provide additional support for projecting parts to relieve these of unnecessary shock Close the carton securely and seal with durable tape Mark the shipping container FRAGILE to ensur...