4.2 PERFORMANCE CHECKS
To determine the amplifier’s performance carry out the following procedure.
4.2.1 Initial Check
The following check can be made after repair and adjustments or
whenever the condition of the unit is in question.
a.
Connect AC power supply. Switch on power and observe that the
display initializes.
b.
Connect a sweep generator (HP 8601 or similar) capable of sweeping
the frequency range 30 KHz to 12 MHz, to the input connector.
c.
Adjust the output level of the sweep generator so that a 50 ohm video
detector connected at the output of the unit will not be damaged by
excessive power output. (Reference section 4.4.1 for set up.)
d.
Observe the gain versus frequency ripple on an oscilloscope calibrated
in decibels. The gain variation must be not more than +/- 1.5 dB over
the frequency range.
e.
Connect a calorimetric power meter (HP435B or equivalent) through a
30 dB 1000 watt attenuator to the output connector. Adjust the input
CW signal to any frequency between 30 kHz and 1 MHz for 500 watts
output.
f.
Observe the harmonic distortion of the output, properly attenuated, on
a spectrum analyzer. The harmonic components contributed by the
amplifier should be better than 25 dB down from the fundamental.
4.3.1 Measurement of Gain
1. Equipment Required (or equivalent):