4.1 INTRODUCTION 
 
The E&I 2100L RF amplifier requires no periodic maintenance. The instrument is 
unconditionally stable and is fail-safe under all load conditions. Damage can only 
be externally caused by the incorrect selection of the AC supply voltage or by an 
input  signal  in  excess  of  the  specified  1  volt  rms  equivalent  to  a  power  level  of 
13dBm. 
 
This  chapter  therefore,  deals  only  with  certain  fundamental  procedures  for  fault 
location. 
 
Performance  limits  quoted  are  for  guidance  only  and  should  not  be  taken  for 
guaranteed  performance  specifications  unless  they  are  also  quoted  in  the 
Specification Section 1.2. 
 
4.2 PERFORMANCE CHECKS 
 
To determine the amplifier’s performance carry out the following procedure. 
 
4.2.1  Initial Check 
 
The    following    check    can    be    made    after    repair    and    adjustments    or 
whenever the condition of the unit is in question. 
 

a. 

Connect  AC  power  supply.  Switch  on  power  and  observe  that  the 
display initializes. 

 

b. 

Connect a sweep generator (HP 8601 or similar) capable of sweeping 
the frequency range 10 KHz to 12 MHz, to the input connector. 

 

c. 

Adjust the output level of the sweep generator so that a 50 ohm video 
detector  connected  at  the  output  of  the  unit  will  not  be  damaged  by 
excessive power output. (Reference section 4.4.1 for set up.) 

 

d. 

Observe the gain versus frequency ripple on an oscilloscope calibrated 
in decibels. The gain variation must be not more than +/- 1.5 dB over 
the frequency range. 

 

e. 

Connect  a  calorimetric  power  meter  (HP434  or  equivalent)  through  a 
10 dB 200 watt attenuator to the output connector. Adjust the input CW 
signal  to  any  frequency  between  10  kHz  and  12  MHz  for  100  watts 
output. 

 

f. 

Observe the harmonic distortion of the output on a spectrum analyzer. 
The  harmonic  components  contributed  by  the  amplifier  should  be 

Содержание 2100L

Страница 1: ...hout notice July 2012 Revision I 1 2100L Broadband Power Amplifier HIGH RF VOLTAGES MAY BE PRESENT AT THE OUTPUT OF THIS UNIT All operating personnel should use extreme caution in handling these volta...

Страница 2: ...ly operational such that it performs according to the original specifications free of charge to the original purchaser Should warranty service be required the unit must be returned to E I freight cost...

Страница 3: ...M FM SSB pulse and any complex modulation signals The amplifier has 50 dB gain it is unconditionally stable and will not oscillate even with combinations of mismatched source and load impedance It is...

Страница 4: ...put load conditions without damage POWER OUTPUT METER True RMS power detection 3 of full scale accuracy from 20 KHz 12 MHz POWER REQUIREMENTS 100 240 VAC 47 63 Hz SIZE 5 3 x 16 5 x 18 4 inches 13 5 x...

Страница 5: ...ms indicated maximum input voltage is 5 times the level of the input signal required to achieve maximum output Input voltages in excess of 2 volts peak may permanently damage the instrument iii Connec...

Страница 6: ...unrestricted If the fault persists please contact Field Service 2 3 3 RS 232 Interface The 2100L features a standard RS 232 serial interface suitable for connection to a PC or host system The communi...

Страница 7: ...l from the input of the driver is coupled to the input of the MMIC front end The output signal of the MMIC is coupled to the gate of transistor Q1 The further amplified signal appearing at the drain o...

Страница 8: ...12 Revision I 8 The main power supply unit provides a 48 VDC 12 ampere source It is a switch mode power supply unit The output feeds the PA and the 24 Volt DC power supply regulator This in turn provi...

Страница 9: ...ocedure 4 2 1 Initial Check The following check can be made after repair and adjustments or whenever the condition of the unit is in question a Connect AC power supply Switch on power and observe that...

Страница 10: ...surement of Gain 1 Equipment Required or equivalent a Osilloscope Tektronix T921 b Sweep Generator HP8601A c Signal Generator Exact Model 7060 d 50 ohm Detector Wavetek D151 e Attenuator 30 dB 200 Wat...

Страница 11: ...el to full deflection Rotate the step attenuator CCW so that the output is reduced by 50 dB g Reconnect the 2100L into the test set up of Figure 4 1 h Place the 2100L power switch to the ON position i...

Страница 12: ...aper or plastic and place in a rigid outer box of wood fiberboard or very strong corrugated cardboard Use ample soft packing to prevent movement Provide additional support for projecting parts to reli...

Отзывы: