background image

 

 

BEFORE: Allows dispensing of medication before the set alarm time. This is 

useful if needing to take a dose earlier than scheduled in order to not miss 

events that take you away from your dispenser. Do not use this function with 

patients who may become confused by its function. This function is turned off 

by default (duration set to 00:00). This setting can be set from 5 minutes to 30 

minutes. 

1.

 

Tap button 2 (decrease) or button 3 (increase) to set the length of the 

BEFORE duration, this turns on the function. 

a.

 

How to Use: If a BEFORE duration has been set, the user may 

access their medication that many minutes before the set dose 

time by turning the dispenser upside down (past 180°) and then 

right side up TWICE in succession. 

b.

 

The dispenser will then rotate, and the next dose will become 

available. If a BEFORE dose is taken, the dispenser will not 

activate again until the next alarm time. 

 

AFTER: Alarm duration. This setting can be set from 5 minutes to 1 hour. 

1.

 

When in menu B, tap button 1 until ‘AFTER’ is displayed. 

2.

 

Tap button 2 (decrease) or 3 (increase) to set the alarm duration. 

SOUND: Alarm melody. 

1.

 

When in menu B, tap button 1 until ‘SOUND’ is displayed.

 

 

2.

 

Tab button 2 (decrease) or 3 (increase) to set the alarm melody. 

a.

 

00 for silent (visual alarm only) and 01 – 04 for sound. 

 

LIGHT: Visual alarm (blinking red light). 

1.

 

When in menu B, tap button 1 until ‘LIGHT’ is displayed. 

2.

 

Tap button 2 to turn off ‘00’ or button 3 to turn on ’01’

 

 

HOURS: Use either a 12-hour clock (AM/PM) or 24-hour clock. 

The dispenser’s default uses the 12-hour clock. 

1.

 

When in menu B, press button 1 until “HOURS” is displayed. 

2.

 

Tap button 2 or 3 to set to: 

a.

 

12 hour – ‘AM’ or ‘PM’ will be displayed next to the time. 

b.

 

24 hour – ‘AM’ or ‘PM’ will not be displayed next to the time. 

 

DATE: This is optional and does not need to be set.  

1.

 

Tap button 1 until ‘DATE’ is displayed. 

2.

 

Wait until ‘YY’ appears. 

3.

 

Tap button 2 (decrease) or 3 (increase) to set the correct YEAR. 

4.

 

Tap button 1 to move to the month ‘MM.’ 

Содержание MedTime 991019

Страница 1: ...1 e pill MedTime Locked Automatic Pill Dispenser Setup Guide Carefully review and save this setup guide Item 991019 ...

Страница 2: ...2 Components Locked Automatic Pill Dispenser x1 Item 991019 Medication Tray x2 Item 991020 Key x2 Item 991022 Dosage Template Set x2 Item 991018 ...

Страница 3: ...ions locked inside even when shaken but if excessive force is applied this action will damage the drive unit inside the dispenser and void the warranty We recommend that you always keep e pill MedTime locked Two keys are included one key for the caregiver and one key as a backup Important This dispenser is locked but it is not tamperproof therefore it should not be used to administer pain medicati...

Страница 4: ... DOSES is displayed 2 Tap button 2 decrease or 3 increase to set the number of alarms ALARMS Set all alarms in order starting with the earliest 1 Tap button 1 until ALARMS is displayed 2 Set the earliest alarm first 1 Tap button 2 to decrease or button 3 to increase the alarm time note AM and PM Tap button 1 to save and continue when the desired alarm time is displayed 3 Repeat for all alarms need...

Страница 5: ...er pieces using a pill splitter can help Begin to Use 1 Place the tray into the dispenser The next dose will be the first compartment to the right of the slot located directly at 6 o clock Important Never force the tray to rotate 2 Close the lid 3 Turn the dispenser over and lock the dispenser WARNING When locking unlocking the cover do not turn the key more than a of a turn Congratulations Your e...

Страница 6: ...ns which has caused damage to the dispenser This is often due to the patient shaking the unit This error message issue occurs when the dispenser has been misused and is therefore not covered under warranty Solution 1 The dispenser will need to be replaced www epill com Item 991019 Error Type 4 TILTED shown on the displayed If the dispenser is lying upside down during the BEFORE and AFTER intervals...

Страница 7: ...til TEST is displayed 4 Tap button 2 to start Test 2 a TURN will appear close the lid and turn the dispenser upside down then back upright TWICE b MOVING will appear on the display The tray will rotate the alarm will sound then TAKE will appear on the display c Close the lid and turn the dispenser upside down to simulate dispensing This will silence the alarm d TEST appearing on the display indica...

Страница 8: ... the next alarm time AFTER Alarm duration This setting can be set from 5 minutes to 1 hour 1 When in menu B tap button 1 until AFTER is displayed 2 Tap button 2 decrease or 3 increase to set the alarm duration SOUND Alarm melody 1 When in menu B tap button 1 until SOUND is displayed 2 Tab button 2 decrease or 3 increase to set the alarm melody a 00 for silent visual alarm only and 01 04 for sound ...

Страница 9: ...us 28 doses were dispensed 1 When in menu B tap button 1 until RECORD is displayed the time at which the last dose was taken is also shown 2 Tap button 2 or 3 to show the previous dispensing times TEST 1 Test rotation function 1 See Error Handling section on pages 6 7 for a solution SAVE and EXIT Menu B 1 Press and hold button 1 for a few seconds to save and exit Menu B 2 The current time will be ...

Страница 10: ... provider This e pill Medication Reminder is only a supplement to any memory system the user currently uses to take medications The user maintains full responsibility to take the correct medication at the appropriate time and to keep the device secure and locked e pill LLC assumes no responsibility for any incidental or consequential damages including without limitation damages or injuries resulti...

Отзывы: