E-Pill MedSmart Скачать руководство пользователя страница 6

 

General Information

 

 

Locking and Unlocking the Dispenser 

WARNING: The 

dispenser’s 

lock should only be turned a ¼ of a turn.  

 

Lock: When closing the lid, you will not hear a click. Please double check that 

the lid is completely closed before turning the dispenser over to lock it. To 

lock, insert the key into the silver lock and turn the key a ¼ of a turn. 

 

Unlock: To unlock, turn the dispenser upside down, insert the key into the 

silver lock and turn the key a ¼ of a turn.  

 

Dispensing 

1.

 

The device will alarm for up to 60 minutes along with blinking lights. 

2.

 

Slide open the white dispensing door if not already open. We 

recommend keeping the dispenser plugged in while dispensing. 

3.

 

Dispense by turning e-pill MedSmart Voice upside down. 

4.

 

Dispensing turns off the alarm. You will be confirmed with 

“Thank 

you.

” 

after medications are dispensed. 

 

Alarm Duration

 

At alarm time a voice r

eminder “It’s time for your medications” will sound for 

up to 60 minutes. The lights will blink every 3 seconds for up to 60 minutes. 

 

Missed Dose

 

If a dose is not dispensed within 60 minutes after the alarm has started, the 

voice reminder message “Patient may have missed their medications” will 

sound every minute for up to 4 minutes. After this, the pills will remain 

available until the next alarm time when the next dose rotates into position, 

and thus rotating away the missed dose. 

 

Low Battery 

We recommend keeping e-

pill MedSmart Voice plugged in as the batteries’ 

main purpose is to supply power to the dispenser in case of a power outage. If 

you choose to run the dispenser on batteries full time, the estimate battery life 

is 6-12 months. When the dispenser starts to run low on battery, the low 

battery icon will appear. At this time, replace all 4 AA batteries as soon as 

possible using only alkaline batteries.  

Important: Only use alkaline/non-rechargeable batteries in e-pill MedSmart 

Voice. The dispenser does not charge batteries. 

Содержание MedSmart

Страница 1: ...1 Instruction Manual e pill MedSmart Voice Locked Automatic Pill Dispenser Item 995040 Carefully review and save this setup guide...

Страница 2: ...ation Tray x1 Item 992040 Locked Automatic Pill Dispenser x1 Item 995040 Dosage Template Set x1 Not Available Separately Key x2 Item 995030 AA Battery x4 Not Available Separately Power Cord x1 Not Ava...

Страница 3: ...that you always keep e pill MedSmart Voice locked Two keys are included one key for the caregiver and one key to be kept as a backup e pill MedSmart Voice is locked but it is not tamperproof therefor...

Страница 4: ...r a click The door is now unlatched Lift up to open Clock Setting the Clock 1 Press and hold button A for a few seconds until stops blinking located between the hours and minutes then release 2 Tap bu...

Страница 5: ...ay easier Important Never force the tray to rotate 2 e pill MedSmart Voice comes with 9 different dosage templates Each template is designed for a specific number of alarms and very importantly the ti...

Страница 6: ...rm time a voice reminder It s time for your medications will sound for up to 60 minutes The lights will blink every 3 seconds for up to 60 minutes Missed Dose If a dose is not dispensed within 60 minu...

Страница 7: ...l Not Silence the Alarm Solutions 1 The batteries are near depletion Replace all 4 AA batteries immediately Only use alkaline 2 If replacing the batteries does not solve the error then the dispenser h...

Страница 8: ...ovider This e pill Medication Reminder is only a supplement to any memory system that you currently use to remind yourself to take medications The user maintains full responsibility to administer medi...