Settings before taking measurements

Before measuring temperature, it is recommended to set the technical 
unit : °C or °F

To set this parameter, open the battery door by pushing on both sides 
of the trigger. It is not necessary to disconnect the battery to make this 
setting.

• 

Set the technical unit

Set the selector unit to °C or °F with 
a screwdriver..

Operating mode

- Press 

ENT

 trigger to turn on the instrument. The backlighted screen, 

indicating the temperature, and the laser turn on. 

- Keep

 ENT

 pressed. Place the laser sighting at the center of the area 

to be measured. 

- Release 

ENT

.

- Read the displayed temperature. (Disaply stays on for 7 seconds 

after the last manipulation).

HOLD

 appears at the top left of the screen ; measurement stays 

displayed.

- Press 

UP

 button ta activate or deactivate the laser ; 

- Press 

DOWN

 button to activate or deactivate the backlight.

- High alarm : when 

KIRAY100

 instrument is turned on, press 

times

 

MODE

 button to set high alarm. 

ON

 or 

OFF

 flashes, press 

UP

 

or

 DOWN

 button to activate or deactivate it (

ON

 or 

OFF

).

Press 

MODE 

button, high alarm temperature is displayed and high 

alarm indicator flashes, press 

UP

 button to increment it or 

DOWN

 button to decrement it.

To return to measurement mode, press 

ENT 

button ; press 

MODE

 

button to go to next mode.

- Low alarm : when

KIRAY100

 instrument is turned on, press 

times

 

MODE

 button to set low alarm.

 ON

 or 

OFF

 flashes, press 

UP

 or 

DOWN 

button to activate or deactivate it (

ON

 or 

OFF

).

Press 

MODE 

button, low alarm temperature is displayed and low 

alarm indicator flashes, press 

UP

 button to increment it or 

DOWN

 

button to decrement it.
To return to measurement mode, press 

ENT 

button ; press 

MODE

 

button to go to next mode.

Emissivity

Emissivity is a term used to describe the energy-emitting 
characteristics of materials.
Most (90% of typical applications) organic materials and painted or 
oxidized surfaces have an emissivity of 0.95 (pre-set in the unit). 
Inaccurate readings will result from measuring shiny or polished 
metal surfaces. To compensate; cover the surface to be measured 
with masking tape or flat black paint. Allow time for the tape to reach 
the same temperature as the material underneath it. Measure the 
temperature of the tape or painted surface.
See table below for values of emissivity of specific materials : 

Important information

For correct measurements :

• Do not take any measurement on metal or shiny or reflective 

surfaces.

• Do not measure through transparent surfaces such as glass, for 

example.

• Water vapor, dust, smoke, etc ... may prevent correct 

measurements because they obstruct the optical of the instrument.

• Make sure that the target is larger than the size of the laser 

sighting.

Command buttons

 Mode button

Allows to define types of desired measurement : emissivity, lock, high 
alarm, low alarm by pressing as many time on this button.

- EMS : when 

KIRAY100

 instrument is turned on, press 

MODE 

button,  

ε

 

flashes. Set emissivity by pressing on 

UP

 button to 

increment it or 

DOWN 

button to decrement it. By default, emissivity is 

set on 0.95.
To back to measurement mode, press 

ENT 

button ;

 

press 

MODE

 

button to go to next mode.

- Lock : when 

KIRAY100

 instrument is turned on, press 

MODE 

button twice, the lock indicator at the top of the screen flashes and

 

OFF

 is displayed. Press 

UP

 or 

DOWN

 button to put the lock 

ON

Press 

MODE

 button to go to next mode, or press 

once ENT

 button : 

KIRAY100

 instrument takes continuous measurements. To cancel 

lock mode, press 

once

 

ENT 

button.

ENT

MODE

Trigger

- Turning on the device.

ENT 

pressed : activation of the laser sighting and temperature 

measurement.
-

 ENT

 released : display is on HOLD (HOLD fixed), and gives the last 

measurement. Display stays on for 7 seconds. If no buttons are 
activated and if continuous measurement is inactivated, the 
instrument turn off after 7 seconds.

Aluminium

0.30

Asbestos

0.95

Asphalt

0.95

Basalt

0.70

Brass

0.50

Brick

0.90

Carbon

0.85

Ceramic

0.95

Concrete

0.95

Copper

0.95

Dirt

0.94

Frozen food

0.90

Hot food

0.93

Glass

0.85

Water

0.93

Ice

0.98

Iron

0.70

Lead

0.50

Limestone

0.98

Oil

0.94

Paint

0.93

Paper

0.95

Plastic

0.95

Rubber

0.95

Sand

0.90

Skin

0.98

Snow

0.90

Steel

0.80

Textile

0.94

Wood

0.94