page 46
Manual
OMNIPORT 20 english
One point calibration of SDI sensors such as combination
probe for humidity / temperature.
CAL1:
CAL1 initiates the one point calibration to channel 1 (Sensor 1).
The symbols for Sensor 1 and Sensor 2 disappear from the
display. The actual value for channel 1 (Sensor 1) is now displayed
in the upper half of the display. The offset value for channel 1
(Sensor 1) flashes in the lower area; this value can be incremented
using and decremented using
. The offset value is confirmed
with .
The menu is than exited automatically and
CAL END
is displayed.
Incorrect calibration is indicated by
CAL FAIL
and must be repeated.
The factory settings can be restored by setting the offset to 0.0.
Important:
Calibration should only be
carried out by skilled personnel using
suitable calibration standards.
Содержание OMNIPORT 20
Страница 33: ...page 33 Manual OMNIPORT 20 english...
Страница 34: ...page 34 Manual OMNIPORT 20 english...
Страница 95: ...page 95 Manuel OMNIPORT 20 fran ais...