TECHNICAL DATA

Air Velocity

Measurement range 

0...5 m/s 

(0...1000 ft/min)

 

 

0...10 m/s 

(0...2000 ft/min)

 

 

0...15 m/s 

(0...3000 ft/min)

 

 

0...20 m/s 

(0...4000 ft/min)

Accuracy

1)

  

±(0.2 m/s / 

40 ft/min

  + 3 % of mv): 

0.5...5 m/s 

(100...1000 ft/min)

 

at 20 °C 

(68 °F) 

/ 45 % RH and 1013 hPa 

(14.7 psi)

 

±(0.3 m/s / 

60 ft/min 

+ 4 % of mv): 

1... 10 m/s 

(200...2000 ft/min)

 

 

±(0.35 m/s / 

70 ft/min 

+ 5 % of mv): 

1... 15 m/s 

(200...3000 ft/min)

 

mv = measured value

 

±(0.4 m/s / 

80 ft/min 

+ 6 % of mv): 

1... 20 m/s 

(200...4000 ft/min)

Analogue output signal  

0 - 1 / 5 / 10 V

2)

,

 

max. 1 mA

Digital interface 

RS485 with Modbus RTU protocol (EE671 = 1 unit load) 

Response time 

τ

90

, typ.  

4 s

General

Supply voltage (Class III) 

 

10 - 29 V DC

Current consumption, max.  

50 mA at 20 m/s 

(4000 ft/min)

 

Connection  

Cable  

0.5 m 

(1.6 ft)

 / 2 m 

(6.6 ft)

 cable, PVC, 5x0.25 mm² (AWG 23) with ferrules 

 

 

 

Plug  

M12 connector, 5-pin

Electromagnetic compatibility

3)

 

EN 61326-1  EN 61326-2-3  Industrial Enviroment

 

 

FCC Part15 Class A  ICES-003 Class A

Material / protection rating 

Polycarbonate / IP50 (probe head); IP54 (enclosure)

Temperature range 

Operation  

-20...60 °C 

(-4...140 °F)

 

 

Storage  

-30...60 °C 

(-22...140 °F)

Humidity range 

5...95 % RH (non-condensing)

1) 

The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-fold standard deviation). The tolerance was calculated in 

 

    accordance with EA-4/02 following the GUM (Guide to the Expression of Uncertainty in Measurement).

2) 0 - 10 V version only with supply voltage ≥ 15 V 

 

3) The EE671 is not short-circuit-proof and not surge-proof (ESD-sensitive device).

SCOPE OF SUPPLY

•  EE671 sensor according to ordering guide

•  Protection cap

•  Mounting flange

•  Quick guide

MODBUS RTU

The EE671 air flow probe represents 1 unit load in a Modbus RTU network. For Modbus protocol settings see Modbus Application 

Note AN0103 (www.epluse.com/ee671). 
Factory settings: Modbus address 238, Baud rate 9 600, even parity, 1 stop bit. The Modbus address can be customised in the 

 

register 0x00 (value margin 1 - 247 permitted). Selectable Baud rates are 9 600, 19 200 and 38 400, parity may be even, odd or none.
Device address, Baud rate and parity can be set via:

1.  PCS10 Product Configuration Software and the appropriate configuration cable HA011018.

The PCS10 can be downloaded free of charge from 

www.epluse.com/pcs

.

2.  Modbus protocol in the register 60001 (0x00) and 60002 (0x01).

Measured value

Unit

Scaling

Type

Register [DEC]

Protocol address [HEX]

Read registers (function code 0x03 / 0x04)

Serial number

ASCII

0001

0x00

Software version

Binary

0009

0x08

Sensor name

ASCII

0010

0x09

Temperature

°C

32 bit float

1003

0026

0x3EA

0x19

Temperature

°F

32 bit float

1005

0028

0x3EC

0x1B

Temperature

K

32 bit float

1009

0030

0x3F0

0x1D

Air velocity

m/s

32 bit float

1041

0032

0x410

0x1F

Air velocity

ft/min

32 bit float

1043

0034

0x412

0x21

Temperature

°C

  x 100  

x 100

16 bit integer

4002

0046

0xFA1

0x2D

Temperature

°F

  x 50  

x 100

16 bit integer

4003

0047

0xFA2

0x2E

Temperature

K

  x 50  

x 100

16 bit integer

4005

0048

0xFA4

0x2F

Air velocity

m/s

  x 100  

x 100

16 bit integer

4021

0049

0xFB4

0x30

Air velocity

ft/min   x 1

 

x 10

16 bit integer

4022

0050

0xFB5

0x31

Write registers (function code 0x06)

Modbus address

0001

0x00

Communication parameters

0002

0x01

1)

2)

3)

1) Please observe correct scaling for used registers

2) Measured values available in both registers 

3) Registers in right column not intended for new design (0026...0050 / 0x19...0x31)

2)

3)

}

 left column E+E standard registers, right column legacy registers