Preliminary operation manual EE600 Differential Pressure Sensor

6  Maintenance and service

EE600 does not require any special maintenance, nevertheless it is recommended to perform a zero 

point adjustment every 12 month. If needed, the enclosure may be cleaned and the device may be 

re-adjusted as described below.

6.1  Cleaning

Use a damp soft cloth to remove deposits of dust or dirt from the exterior of the device and from the 

display. Do not attempt to clean the interior of the device. Do not use any solvents, alcohol or abrasive 

cleaning agents.

6.2  Repairs

Repairs may be carried out by the manufacturer only. The attempt of unauthorized repair excludes any 

warranty claims. 

6.3  Readjustment of EE600

A periodical readjustment of EE600 might be required by the regulations of certain industries or by the 

need of best long-term measurement accuracy.  The zero point and the span point can be adjusted with 

push buttons on the EE600 electronics board. 

For adjustment the device must be powered and the enclosure cover removed. Consequently, the 

adjustment may be performed by authorized staff only, observing the handling of electrical sensitive 

devices (ESD). 

6.3.1  Zero Point Adjustment

The zero point adjustment is used to correct an eventual zero point deviation. 

a.  Remove the tubes from both pressure connections of the EE600. By this the pressure is same at 

both connections. 

b.  Press “zero point” button 1-2sec. (see „3 Product Description“). 

c. 

The successful zero point adjustment is confirmed by the green LED turning on for 2 seconds. 

d.  The red LED turning on for 2 seconds indicates that the zero point adjustment did not succeed. In 

this case repeat b.

e.  The zero point adjustment (b.) may not succeed because of a deviation higher than 5% of the  

original full scale of the device. This is 50 Pa for the 0 – 1000 Pa range and 500 Pa from the 0 – 

10000 Pa range. In such a case, the zero point adjustment can be forced by pressing and holding 

the zero point button for min. 10 sec. 

f. 

The successful zero point adjustment is confirmed by the green LED turning on for 2 seconds. 

6.3.2  Span Point Adjustment

The span point adjustment is used to correct an eventual deviation of the span value. 

Important: 

Make sure to perform a zero point adjustment as above before any span point adjustment. 

a.  Connect the differential pressure reference device/calibrator to the EE600 and set it according to 

the EE600 span value to be adjusted.

Observe the operation manual of the reference device.

b.  Press the span point adjustment button (see „3 Product Description“). 

c. 

The successful span point adjustment is confirmed by the green LED turning on for 2 seconds. 

Содержание EE600

Страница 1: ...BA_EE600_e v1 1 technical data are subject to change Operation Manual EE600 Differential Pressure Sensor...

Страница 2: ...and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the...

Страница 3: ...e with S3 and S4 7 4 3 Select the display unit with S5 and S6 7 4 4 Select the backlight for the display 7 4 5 Select the Output Signal with S8 7 5 Installation 8 5 1 Pressure connection 8 6 Maintenan...

Страница 4: ...damage the device Do not use EE600 in explosive atmosphere or for measurement of aggressive gases Do not apply the nominal voltage to the output terminals 1 1 2 Mounting start up and operation The EE6...

Страница 5: ...ng conditions Consequential damages are excluded from the liability 1 2 Environmental aspects Products from E E Elektronik are developed and manufactured observing of all relevant requirements with re...

Страница 6: ...LED Fig 2 Design and components 15 35 V DC 24 V AC 20 Fig 3 Connection diagram 4 Initial Setup Each EE600 leaves the E E factory with the default setup all switches on 0 Measurement range 100 of the f...

Страница 7: ...e response time with S3 and S4 Tab 2 DIP switch settings Response time 4 3 Select the display unit with S5 and S6 Tab 3 DIP switch settings Display unit 4 4 Select the backlight for the display S7 Bac...

Страница 8: ...3 Connection diagram Close tightly the cable gland This is essential for the compliance with the IP65 NEMA4 protection class of the EE600 enclosure Fig 5 Mounting orientation 5 1 Pressure connection F...

Страница 9: ...ose route Important Make sure to connect the higher pressure at the pressure connector and the lower pressure at the pressure connector Inverted connection leads to out of range information on the opt...

Страница 10: ...is used to correct an eventual zero point deviation a Remove the tubes from both pressure connections of the EE600 By this the pressure is same at both connections b Press zero point button 1 2sec see...

Страница 11: ...a case the set point adjustment can be forced by pressing and holding the span point button for 10 sec f The successful zero point adjustment is confirmed by the green LED turning on for 2 seconds 6 3...

Страница 12: ...r negative pressure due to inverse pressure connection the display indicates out of range out of range p 5 MR or inverse pressure connection p 105 MR Fig 10 Failure indication Green LED flashing 1s in...

Страница 13: ...5 Additionally the analogue outputs indicate measurement out of range as follows Indication of p 5 MR and of negative difference pressure due to inverted pressure connection Output signal Indication...

Страница 14: ...1 mA selectable with and4 DIP switches1 0 20 mA or 4 20 mA 3 wire RL 500 Ohm General Power supply 15 35 V DC or 24 V AC 20 Current consumption typ 18 mA without Display at 0 Pa 24 V DC typ 43 mA with...

Страница 15: ...15 Preliminary operation manual EE600 Differential Pressure Sensor...

Страница 16: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Отзывы: