User Manual
EE381
Page 13
•
Display shows incorrect values:
ꟷ Error during re-adjustment of the sensor
Reset to factory calibration and
repeat the calibration routine
ꟷ Output configured incorrectly
Check output range and output signals
in the configuration
•
Sensor failure:
ꟷ No supply voltage
Check wiring and supply voltage
Green LED is illuminated continuously: electronics defect,
contact the manufacturer
Further self-diagnosis:
i
•
Error:
ꟷ possible cause
measures / help
6.3 Replacement Parts / Accessories
Description
Order code
Stainless steel filter for flow > 1 m/s
HA010110
Display
D08-EE381
Product Configuration Adapter see datasheet
(See also the EE-PCA datasheet)
EE-PCA
E+E Product Configuration Software
(Free download: www.epluse.com/configurator)
EE-PCS