![E+E Elektronik EE23 series Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/e-e-elektronik/ee23-series/ee23-series_user-manual_3071037023.webp)
23
TECHNICAL DATA
Measured quantities
Relative humidity
Humidity sensor
1)
EE23-xA/B/Cx
HC1000-200
EE23-xHx
HC105
Working range
1)
0...100% RH
Accuracy
2)
(including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV...)
EE23-xA/B/Cx EE23-xHx
-15...40°C
(5...104°F)
≤
90% RH
± (1.3 + 0.3%*mv) % RH
± (1.8 + 0,3%*mv) % RH
-15...40°C
(5...104°F)
>90% RH
± 2.3% RH
± 2.8% RH
-25...70°C
(-13...158°F)
± (1.4 + 1%*mv) % RH
± (1.9 + 1%*mv) % RH
-40...120°C
(-40...248°F)
± (1.5 + 1.5%*mv) % RH
-
Temperature dependence electronics
typ. ± 0.015% RH/°C
Response time with metal grid filter at 20°C / t
90
< 15 sec.
Temperature
Temperature sensor element
EE23-xA/B/Cx
Pt1000 (class A, DIN EN 60751)
EE23-xHx
Pt1000 (class B, DIN EN 60751)
Working range sensing head EE23-xAx -40...60°C
(-40...140°F)
EE23-xBx -40...80°C
(-40...176°F)
EE23-xCx -40...120°C
(-40...248°F)
EE23-xHx -40...80°C
(-40...176°F)
Accuracy
Temperature dependence of electronics
typ. 0.002°C/°C
Outputs
0 - 1 V
-0.5 mA < I
L
< 0.5 mA
0...100% RH / xx...yy°C
3)
0 - 5 V
-1 mA < I
L
< 1 mA
(temperature output scale adjustable by E+E or
0 - 10 V
-1 mA < I
L
< 1 mA
with configuration kit)
0 - 20mA
R
L
< 470 Ohm
4 - 20 mA
R
L
< 470 Ohm
Max. adjustable output scaling
4)
from
up to
units
EE23-A
EE23-B, H
EE23-C
Humidity
RH
0
100
100
100
% RH
Temperature
T
-40
(-40)
60
(140)
80
(176)
120
(248)
°C
(°F)
Dew-point temperature
Td
-40
(-40)
60
(140)
80
(176)
100
(212)
°C
(°F)
Frost-point temperature
Tf
-40
(-40)
0
(32)
0
(32)
0
(32)
°C
(°F)
General
Supply voltage
for 0 -1 V, 0 - 5 V outputs
10.5 - 35V DC or 12 - 28V AC
for 0 - 10 V, 0 - 20 mA and 4-20 mA outputs
15.0 - 35V DC or 15 - 28V AC
(optional 100...240V AC, 50/60Hz)
Current consumption for voltage output
for DC supply
≤
25 mA
with alarm module: for DC supply
≤
35 mA
for AC supply
≤
35 mA
eff
for AC supply
≤
60 mA
eff
Current consumption for current output
for DC supply
≤
50 mA
with alarm module: for DC supply
≤
60 mA
for AC supply
≤
90 mA
eff
for
AC
supply
≤
110 mA
eff
Housing / protection class
PC or Al Si 9 Cu 3 / IP65;
Nema 4
Cable gland
5)
M16x1.5
cable Ø 4.5 - 10 mm
(0.18 - 0.39”)
Electrical connection
5)
screw terminals max. 1.5 mm²
(AWG 16)
Working temperature range of electronics
-40...60°C
(-40...140°F)
Working temperature range with display
-30...60°C
(-22...140°F)
Storage temperature range
-40...60°C
(-40...140°F)
CE compatibility according
EN61326-1
EN61326-2-3
ICES-003 ClassB
Industrial Environment
FCC Part15 ClassB
Alarm Module - optional
Output
SPDT-Switch up to 250V AC/8A or 28V DC/8A
threshold
hysteresis
Setting range
10...95% RH
3...15% RH
Setting accuracy
± 3% RH
1) Refer to the working range of the humidity sensor
3) Refer to ordering guide
4) Refer to accuracies of calculated values
5) Connection plugs refer to ordering guide
2) The accuracy statement includes the uncertainty of the factory calibration with an enhancement
factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02
and with regard to GUM (Guide to the Expression of Uncertainty in Measurement)
°C
EE23-xA/B/Cx
EE23-xHx
°C
∆°
C
∆°
C
Schaltpunkt
switching point
[%]
Relais
S
tatus
relay
st
atus
70
62
Ausschaltpunkt
switching off
Einschaltpunkt
switching on
EIN
ON
AUS
OFF
8%
NC
C
NO
NC
C
NO
HYSTERESIS
9
7
5
4
3
11
15