User Manual for EE08 Humidity / Temperature Probe

EE08 with connector:

EE08-MxE11xxx

T active / passive, 2 wire

Assignment of M12 connection cable 

(HA010322, HA010323, HA010324, HA010325)  

1 T passive

1)

white

2 SDA

}

 

E2-

interface

brown

3 SCL

green

4 RH out

yellow

5 T out

grey

6 GND

pink

7 T passive

1)

blue

8 +UB

red

1) not connected for T active versions (M1)

Ground connection:

A low impedance connection between the shield of the connection cable and the ground potential is 

important for the flawless operation of the EE08.  

4  Installation

The follwing mounting types are possible:

 

ƒ

Wall mount with the help of a mounting clip, available as accessory HA010211.

 

ƒ

Outdoor operation with radiation shield: wall mount or pole mount. Please mind the mounting 

instructions included in the manuals of HA010502 and HA010506.

5  Maintenance

The use in dirty, dusty, polluted environment might arise the need for cleaning the sensing head and 

replacing the filter cap. In such a case please see the Cleaning Instructions at www.epluse.com/EE08. 

Do not touch the humidity sensor! 

6  Calibration / Adjustment

Definitions

 

ƒ

Calibration

 documents the accuracy of a measurement device. The device under test (specimen) is 

compared with the reference and the deviations are documented in a calibration certificate. During 

the calibration, the specimen is not changed or improved in any way.

 

ƒ

Adjustment

 improves the measurement accuracy of a device. The specimen is compared with the 

reference and brought in line with it. An adjustment can be followed by a calibration which documents 

the accuracy of the adjusted specimen.

To carry out a one point or a two point calibration / adjustment, the E2 / RS232 converter (available as 

an accessory, order code HA011005) and the EE-PCS Product Configuration Software are necessary. 

The EE-PCS is freely available at www.epluse.com/EE08.

Содержание EE08

Страница 1: ...BA_EE08 v1 3 Modification rights reserved EE08 High Precision Miniature Humidity and Temperature Probe User Manual...

Страница 2: ...es and can radiate high frequency energy If it is not installed and used in accordance with the operating instructions it may cause electromagnetic interference to radio communications However there i...

Страница 3: ...ons 4 1 2 2 Intended Use 4 1 2 3 Mounting Start up and Operation 5 1 3 Environmental Aspects 5 2 Scope of Supply 5 3 Product Description 5 3 1 General 5 3 2 Dimensions 6 3 3 Electrical Connection 6 4...

Страница 4: ...ing the filter cap make sure not to touch the sensing elements The device must be operated with the filter cap on at all times For sensor cleaning please see Cleaning Instructions at www epluse com EE...

Страница 5: ...into operation Do not install or start up a device supposed to be faulty Make sure that such devices are not accidentally used by marking them clearly as faulty A faulty device may only be investigate...

Страница 6: ...xxx EE08 with cable Type E8 EE08 MxE8xxx M12x1 61 2 4 73 2 87 25 1 12 0 47 connector M12x1 8 poles M16x1 5 61 2 4 71 2 8 83 3 27 12 0 47 Fig 1 Dimensions of EE08 in mm inch 5 1 5 20 5 0 2 20 0 8 122 4...

Страница 7: ...e The use in dirty dusty polluted environment might arise the need for cleaning the sensing head and replacing the filter cap In such a case please see the Cleaning Instructions at www epluse com EE08...

Страница 8: ...Accuracy at nominal voltage1 Outputs Analogue 0 1 V 0 2 5 V 0 5 V 0 10 V 0 2mA IL 0 2 mA Digital interface E2 interface2 General Supply voltage for output 0 1 V 0 2 5 V V1 4 5 15 V DC V2 7 30 V DC for...

Страница 9: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Отзывы: