background image

Proper insertion depth is necessary to achieve maximum ambient noise
attenuation and interaural attenuation. The graph below depicts interaural
attenuation as a function of insertion depth.

The described interaural

attenuation is obtained for each ear in which the E-A-RLINK

®

is fully inserted,

regardless of whether it is the “sending” ear or “receiving” ear.

2. Background Noise Reduction
A major benefit of the E-A-RTONE

®

Insert Earphone in combination with the

E-A-RLINK

®

disposable foam eartips is the reduction of background noise that

might interact with the presented pure tones and influence threshold
determinations. When E-A-RLINK

®

eartips are fitted to the subject according

to the instructions, the expected noise attenuation will typically exceed 30 dB
in the frequency region 125-8000 Hz. This additional attenuation will improve
test reliability in non-ideal environments and, in most cases, allow testing
down to 0 dB HL in a background noise level not exceeding 45 dB(A).

The current ANSI Standard (ANSI S3.1-1999)  “Maximum Permissible
Ambient Noise Levels (MPANLs) For Audiometric Test Rooms” (Table 1
and Table 2) includes octave band and one-third octave band permissible
noise levels for both supra-aural and Insert Earphones. The ANSI octave
band MPANLs are presented below for standard earphones and also for
E-A-RTONES

®

with fully inserted E-A-RLINKS

®

.

4

-10

0

10

20

30

40

50

60

70

80

90

100

110

dB

125

250

500

1000

2000

4000

8000

Hz

Obtained Interaural Attenuation (E-A-RTONE 3A Data)

Full E-A-RLINK insertion in both ears

Shallow insertion in dead ear only

Shallow insertion in both ears

Содержание E-A-RTONE 5A

Страница 1: ...Instructions for the use of Insert Earphones Auditory Systems ...

Страница 2: ...ct coupling also produces much greater sound separation between ears across the test frequency range Clinically that means more instances where the need for masking is eliminated or the level necessary for effective masking is lower Your average testing time will decrease without sacrificing reliability The method of clinical masking you use does not need to be altered Simply substitute the greate...

Страница 3: ...between an insert earphone and a supra aural earphone This instruction booklet is intended to provide information to allow you to perform hearing testing with your audiometer and E A RTONE 5A Insert Earphones with the confidence that test results will be comparable to or in certain instances more reliable than measurements performed with supra aural headphones Most questions regarding the use and ...

Страница 4: ...in extremely small ear canals In these instances the newly available Infant Eartips are appropriate Secure the section of black tubing which protrudes from the E A RLINK eartip to the sound tube nipple protruding from the E A RTONE 5A case Slowly roll the E A RLINK tip into the smallest diameter possible and insert the E A RLINK well into the ear canal see illustrations The case will function as a...

Страница 5: ...ons the expected noise attenuation will typically exceed 30 dB in the frequency region 125 8000 Hz This additional attenuation will improve test reliability in non ideal environments and in most cases allow testing down to 0 dB HL in a background noise level not exceeding 45 dB A The current ANSI Standard ANSI S3 1 1999 Maximum Permissible Ambient Noise Levels MPANLs For Audiometric Test Rooms Tab...

Страница 6: ...insertion of the connected E A RLINK will place the case portion in front of the pinna Always use E A RLINK tips in both ears or block the non test ear with an E A R foam plug 4 Cable Assembly The junction or pod portion of the cable assembly should be supported at or near the patient s mid chest level to relieve tension on the E A RLINK couplers This can be accomplished by attaching the Velcro ba...

Страница 7: ...the microphone a 2cc coupler e g Bruel Kjaer DB 0138 is attached The white nylon tube nipple of the E A RTONE 5A earphone is connected directly to the coupler see illustration Set the intensity dial on the audiometer to 70 dBHL and adjust the output from the audiometer to the SPL values given for the different test frequencies in the table below Repeat the procedure for all available test frequenc...

Страница 8: ... calibrated for a supra aural earphone and the Target SPLs are the audiometer HL setting plus the insert earphone RETSPLs for the particular 2cc coupler used The process will work as well in reverse i e audiometer calibrated for an E A RTONE insert earphone with a supra aural earphone as the secondary transducer as long as the appropriate Target SPLs TDH values are employed Calibrate the audiomete...

Страница 9: ...its for E A RTONE 5A Impedance 10 Ohm or 50 Ohm 1 kHz Sensitivity 102 5 dB SPL in DB 0138 HA 2 coupler 0 1 Volt rms 10 Ohm 0 2 Volt rms 50 Ohm Limits 3 dB Distortion 3 measured 500 Hz 118 5 dB SPL Maximum output Meets or exceeds 110 dBHL at standard audiometric frequencies between 500 and 4000 Hz Safe operating limits Max continuous sine wave drive 2 5 Volt rms 10 Ohm 5 0 Volt rms 50 Ohm Max Peak ...

Страница 10: ... s of the product original date of purchase and the nature of the problem Merchandise not returned prepaid will be refused Mailing at the time of purchase a completed Warranty Card that is included with each E A RTONE Insert Earphone will facilitate our ability to return repaired units as quickly as possible Warranty service will vary depending on the availability of parts and the workload There w...

Страница 11: ...rticular purpose and merchantability and there are no warranties that extend beyond the description or duration of this warranty No representative agent dealer or employee of Aearo Company is authorized to give any other warranty or to assume for Aearo Company any other liability in connection with sale or service of the products covered by this limited warranty In order for Auditory Systems to pr...

Страница 12: ...free from the US only 800 624 5955 Fax 877 807 7113 International Customer Service Ph 317 656 5713 Fax 317 656 5760 Technical Support Ph 317 692 6550 Fax 317 692 6604 2000 Aearo Company Auditory Systems E A R E A RTONE and E A RLINK are all registered trademarks of Aearo Company 018 9013 ...

Отзывы: