 OPERATING GUIDE 

 

MODEL 3196C HIGH TEMPERATURE CHARGE MODE ACCELEROMETER 

INTRODUCTION 
 
 

Model 3196C is a charge mode accelerometer 

designed to measure vibration of surfaces at 
temperatures up to +500 degrees F. This accel- 
erometer uses an ultra stable piezoceramic crystal 
material in its self-generating seismic element. This 
material exhibits less than +9% increase in sensitivity 
from room temperature to +500 degrees F. 
 

Model 3196C may be used with various 

charge amplifiers of the vibration type. (As opposed to 
the DC coupled electrostatic types.) 
 

We especially recommend the Dytran in-line 

charge amplifier Model Series 4705A, for use with the 
3196C. This inexpensive miniature charge amplifier 
adapts the 3196C for use with any LIVM (constant 
current type) power unit. Model 4705A features BNC 
input and output connectors and mounts in-line 
between the 3196C and the power unit. 
 

DESCRIPTION 

 
Refer to Outline/Installation drawing 127-3196C. 
 
 

Model 3196C is packaged in a hermetically 

sealed 300 series stainless steel housing and features 
a transverse mounted ceramic-to-metal sealed TNC 
connector. The unit is installed by use of 3 8-32 x 1/2 
inch long socket head cap screws, (supplied). The 
mounting screws thread into 3 matching threaded 
holes which must be provided in the mounting surface. 
 

Model 3196C also features a connector 

housing design which will accept Dytran's "Immersion 
Proof" cable/boot assembly to protect the cable 
connection against moisture and other contaminants. 
 

INSTALLATION 

 
 

To mount Model 3196C, it is necessary to drill 

and tap 3 equally spaced 8-32 mounting ports on a 
1.188 dia. bolt circle. For best high frequency 
response, the contact area of the accelerometer must 
be selected or prepared to be flat to .001 TIR. The 
holes  must be drilled perpendicular to the mounting 
surface to within 2 degrees of angular error.  
 

After drilling and tapping, clean the area to 

remove all traces of cutting oil and machining chips. 
 

Spread a thin layer of silicone grease on the 

three contact surfaces of the 3196C. Locate the accel. 
over the tapped holes and engage the three mounting 
screws through the holes in the flange and into the 
tapped holes. Thread the mounting screws into the 
tapped holes and torque to 15 pound inches. 

 

Inspect the mating surfaces to ensure that the 

accelerometer is snugged down tightly in intimate 
contact with the test surface at all three contact points. 
 

If the surfaces appear to be meeting squarely, 

the accelerometer is ready for connection of the cable. 
 

INTERCONNECTION 

 
 

The recommended cable assy for the 3196C 

is the Dytran Model 6425A03. This cable is a 3 feet 
long low-noise coaxial with TNC connector on both 
ends.  
 

Connect the cable to the 3196C and connect 

the other end of the cable to the charge amplifier. 
Apply lock wire if applicable.  
 

If you are using a fixed sensitivity charge 

amplifier such as the Model 4705A, the system 
sensitivity is the product of the charge sensitivity of the 
accel. (in pC/G) and the sensitivity of the charge 
amplifier (in mV/pC). 
 
 

Example: 

 
 

Accelerometer sensitivity: 8.0 pC/G. 

 

Charge amplifier (Model 4705M2) sensitivity: 

 

12.5 mV/pC. 

 
 

System sensitivity is: 

 
 

8.0 pC/G x 12.5 mV/pC = 100 mV/g 

 

MAINTENANCE AND REPAIR

 

 
 

Should you experience a problem with your 

system, contact the Dytran factory for technical 
assistance in analyzing and trouble shooting the 
problem. If the product must be returned for evaluation 
and/or repair, you will be given an RMA (returned 
materials authorization) number and instructions for 
returning the instrument to the factory. Do not return 
the instrument without first obtaining this authorization 
to return. 