Referring to Figure 1, the feedback resistor Rf 

in conjunction with shunt capacitance Cf, forms a first 
order high-pass filter which sets the low frequency 
response of the accelerometer in accordance with the 
following equation: 
 
 

 

      .16 

 

f-3db  =  ------------------          (eq.1) 

 

 

       RC 

 
 where: 
 
 

f-3db  =  lower -3db frequency (Hz) 

 
 

R  =  resistance value R   (Ohms) 

 
 

C  =  total shunt capacitance C  (Farads) 

 
 

RC  =  discharge time constant TC (Seconds) 

 
 

Equation 1 above, defines the frequency at 

which the accelerometer sensitivity will be 3db down 
when compared to the reference sensitivity measured 
at 100 Hz. 
 
 

The discharge time constant for all Models is 

0.5 seconds, yielding a lower -3db frequency of 0.3 
Hz, from equation 1. 
 
 

As rule of thumb, the lower -5% frequency is 

three times the -3db frequency or 1 Hz. 
 

INSTALLATION 

 
 

(Refer to Outline/Installation drawing 127-

3049E) To install Model 3049E series accelerometers, 
it is necessary to prepare (or find) a flat mounting area 
of approximately 0.5 inch diameter. Ideally, the 
mounting surface should be flat to .001 in. TIR. The 
flat mounting surface ensures intimate contact 
between accelerometer base and mounting surface for 
best high frequency transmissibility, thus accuracy. 
 

To install the adhesive mounted 3049E1 or 

3049E3, clean the mounting surfaces with solvent if 
necessary to remove all traces of oils and other 
impurities including burrs or any matter which could 
preclude intimate contact between mating surfaces. 
 
 

Apply a light coating of cynoacrylate adhesive 

(or other type of suitable adhesive) to either mounting 
surface, position the accelerometer in the desired 
cable orientation and press the accelerometer firmly 

onto the mounting surface and hold for several 
seconds. 
 
 
 

To install the stud mounted 3049E or 3049E2 

accelerometers, drill and tap a 10-32 mounting hole 
with enough thread depth to accommodate the integral 
stud. Clean the area to remove all traces of machining 
chips, burrs, etc.  Spread a light coating of silicone 
grease, or other lubricant, on either of the mating 
surfaces and thread the accelerometer/stud 
combination into the tapped hole by hand, until the 
accelerometer base seats against the mounting 
surface. Check to see that the mating surfaces are 
meeting properly, i.e., that they are meeting flush and 
that there is not an angle formed between the two 
surfaces indicating that they are not co-planar. If this 
condition is observed, torquing the accelerometer 
down will strain the base causing possible poor 
frequency response and even erroneous reference 
sensitivity. Inspect the perpendicularity of the tapped 
hole.  If the hand tight meeting between the two 
surfaces is satisfactory, torque the 3049E or 3049E2 
accelerometer to the mating surface with 15 to 20 lb-
inch of torque, preferably measuring the torque with a 
torque wrench torquing on the hex surface only. 
 
 

Proper torque will ensure the best high 

frequency performance from the instrument as well as 
repeatability of sensitivity when mounting and 
remounting. Excessive torque could damage the 
ground isolation base. 
 
 

Connect the cable (typically Models 6010AXX 

or 6011AXX) to the accelerometer snugging up the 
threaded lock ring tightly by hand. 

 

NOTE:

 Do not use a pliers or vise grips on the 

knurled lock ring. This could damage the 
connector of the 3049E Series and/or the 
cable connector. 

 
 

To avoid stressing the cables which could lead 

to early failure, especially under larger excursions of 
the test object, it is good practice to tie the cable down 
to a fixed surface near the mounting area at a point 
approximately one inch from the accelerometer. 
 
 

If there is excessive motion between the 

accelerometer and the nearest tie point, allow a strain 
loop of cable to let relative motion occur without 
stressing the cable. 