Chapter 4 Operating the Agility System
78
Agility System
Operator’s Manual
4.10 Aborting Plates/Stopping the System/Shutting the
System Down
4.10.1
Aborting Plates/Stopping the System
The sequence to Abort a Plate or Stop the System (Pause) is initiated by the
button which presents the Pause/Stop/Abort screen (Figure 4-32).
Figure 4-32: Pause/Stop/Abort Screen
The
Plate Abort
option
stops the processing of current plate(s) that have
failed or all plates, depending on the user selection. All unused plates and SmartKits used for
to those plates will be returned to the loading bays. The system may not stop immediately.
Some actions being performed at the time may need to arrive at a state where the system can
be safely stopped before discontinuation. The plate may continue through the operation
sequence but no further samples will have been delivered to it subsequent to plate aborting.
Plate maps will be generated for aborted plates automatically. Plate maps will be located in a
time and date stamped subdirectory in the C:\Program Data\Dynex Technologies\Plate folder.
Содержание Agility Automated ELISA System
Страница 6: ...Table of Contents 6 Agility System Operator s Manual This page intentionally left blank ...
Страница 8: ...About this Manual 8 Agility System Operator s Manual This page intentionally left blank ...
Страница 20: ...Chapter 1 Overview 20 Agility System Operator s Manual Figure 1 17 USB Hub Setup Figure 1 18 USB Hub Setup ...
Страница 24: ...Chapter 2 Safety 24 Agility System Operator s Manual This page intentionally left blank ...
Страница 42: ...Chapter 3 System Design 42 Agility System Operator s Manual This page intentionally left blank ...
Страница 52: ...Chapter 4 Operating the Agility System 52 Agility System Operator s Manual Figure 4 9 Wash Assay Screen ...
Страница 74: ...Chapter 4 Operating the Agility System 74 Agility System Operator s Manual Figure 4 28 Consumables Status ...
Страница 82: ...Chapter 4 Operating the Agility System 82 Agility System Operator s Manual This page intentionally left blank ...
Страница 98: ...Chapter 6 The Graphical User Interface 98 Agility System Operator s Manual Figure 6 1 The Administrator Screen ...
Страница 101: ...Chapter 6 The Graphical User Interface Agility System Operator s Manual 101 Figure 6 4 The Version Report Screen ...
Страница 144: ...Chapter 7Error Handling 144 Agility System Operator s Manual This page intentionally left blank ...
Страница 148: ...Appendix A Specifications 148 Agility System Operator s Manual This page intentionally left blank ...
Страница 150: ...Appendix B Consumables and Spare Parts 150 Agility System Operator s Manual This page intentionally left blank ...