8 Testing
8.2 Testing the DX2-RJM / DX2-RJM-LF
The installation must be fully tested after all modules and cables have been installed. The testing
procedure is detailed in the DX System Manual (section 8 Testing), which must be read in conjunction
with this and other DX/DX2 module manuals.
Warning
:
Do not connect the battery positive (B+) terminal of the battery to the DX2 system until the wheelchair is lifted off
the ground. To prevent the risk of injury, Dynamic Controls recommends the use of a lifting device when lifting the
wheelchair off the ground.
8.1 Before testing
Check that all cables and modules in the system are connected correctly. Check especially that the
polarities of the batteries, the motors and the park brakes are connected correctly and that the
polarities are not swapped.
To prevent the wheelchair from suddenly driving away when you turn it on, put blocks under the
wheelchair frame to lift the wheels off the ground. Check that the wheels can turn freely.
Make the final connection to the battery positive (B+) terminal and close the circuit breakers.
Turn on the DX2 system with the power button on the master remote module and program the DX2
system for the appropriate wheelchair application.
Turn off the DX2 system with the power button.
8.2 Testing the DX2-RJM / DX2-RJM-LF
8.2.1 Power button
l
Test that the system powers up and down correctly with the power button. For more infor-
mation on the power button, see
7.1.2 Power button and status indicator
GBK65701 DX2 Compact Remote Modules
Installation Manual Issue 1
Test - Page 37