background image

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ACCU-CAL

 50-LED

 

LED Radiometer User Guide

 

 

Содержание ACCU-CAL 50-LED

Страница 1: ...ACCU CAL 50 LED LED Radiometer User Guide ...

Страница 2: ...pection and increase throughput System designs enable stand alone configuration or integration into your existing assembly line Please note that most dispensing and curing system applications are unique Dymax does not warrant the fitness of the product for the intended application Any warranty applicable to the product its application and use is strictly limited to that contained in the Dymax stan...

Страница 3: ...t Overview 5 Installation 6 Unpacking and Inspecting Your Shipment 6 Parts List Spot Line Array Configuration 6 Parts List Flood Configuration 6 Operation 7 Testing a Flood Lamp System 7 Testing Spot Lamp Systems 8 Testing the BlueWave QX4 10 Maintenance 12 Spare Parts and Accessories 13 Specifications 13 ...

Страница 4: ...ts technique ou service client à 860 482 1010 Dymax formé professionals attendre a vous servir Safety General Safety Before continuing please read the following chapters of this manual for safety recommendations and installation running and troubleshooting instructions CAUTION Always wear protective goggles or a face shield when working near the front of any unit which emits UV light The rear of s...

Страница 5: ...The ACCU CAL 50 LED Radiometer is a microprocessor based measurement instrument designed to measure curing energy radiation in the range of 350 450 nm The ACCU CAL 50 LED uses two AA batteries Environmental Considerations Optical measurement instruments are sensitive to extremes in environmental conditions like high temperature humidity and contamination Protect the device and its detector s from ...

Страница 6: ...those listed below If parts are missing contact your local Dymax representative or Dymax Customer Support to resolve the problem Figure 2 ACCU CAL 50 LED Radiometer PN 40505 for Spot Line Flood Lamps Figure 3 ACCU CAL 50 LED Radiometer PN 40519 for Flood Lamps Only Parts List Spot Line Array Configuration ACCU CAL 50 LED Radiometer ACCU CAL 50 LED Radiometer User Guide 3 mm 5 mm 8 mm Lightguide Ad...

Страница 7: ...e Light Source button to select the light source being measured Repeatedly press the button until Flood Lamp appears on the top line of the display Figure 5 5 Press and release the Mode button to select the operating mode Repeatedly press the Mode button until the mode you desire appears on the second line of the display The different operating modes are Peak Intensity Shows the highest intensity ...

Страница 8: ... and the second line displays the mode 9 Press and release the Light Source button to select the light source being measured For spot lamps use the lightguide setting Repeatedly press the Light Source button until the correct lightguide size appears on the top line The options are 3 mm Lightguide For use when measuring energy at the end of a 3 mm lightguide 5 mm Lightguide For use when measuring e...

Страница 9: ...Mode button to select the operating mode Repeatedly press the Mode button until the mode you desire appears on the second line of the display The different operating modes are Peak Intensity Shows the highest intensity in mW cm2 seen by the detector during the measurement and stores that value in the display Intensity Shows the intensity in mW cm2 at the detector during the measurement This is a d...

Страница 10: ...sensor to energy emissions for more than 15 seconds Testing the BlueWave QX4 1 Attach the detector to the radiometer Figure 15 2 Attach the BlueWave QX4 optic adapter to the detector Figure 16 using the two 2 mm screws provided Note The optical adapter and optometer are calibrated as a set so they are not interchangeable among other similar type radiometers 3 Insert the LED head into the optic ada...

Страница 11: ...d 8 mm Spot For use when measuring a BlueWave QX4 LED head with the 8 mm focus lens installed Figure 18 Spot Mode Screens Shown in Peak Intensity Mode 7 Press and release the Mode button to select the operating mode The different operating modes are Peak Intensity Shows the highest intensity in mW cm2 seen by the detector during the measurement and stores that value in the display Intensity Shows ...

Страница 12: ...service is available through Dymax Customer Service or Dymax Product Repair Change the batteries when a low battery warning is received The ACCU CAL 50 LED uses two AA type batteries The battery compartment is on the back of the instrument Figure 19 Battery Compartment Closed Open Keep the detector head s sensing element clean and free of contaminants The detector head may be cleaned with a clean ...

Страница 13: ...558 5 mm Lightguide Adapter 39557 3 mm Lightguide Adapter 39556 Line Array Adapter 43364 2 mm Hex Wrench 38656 2 mm Screws 38881 Specifications Property Specification Power Two AA size batteries Display LCD Graphic Display 97x 32 Pixel Display area 0 56 in x 1 41 in 14 3 mm x 35 8 mm Spectral Range 350 450 nm Detector Interface 9 Pin MDSM9 socket 4 inputs Measurement Ranges Four modes of operation...

Страница 14: ...nt may void your warranty benefits The use of aftermarket replacement parts not supplied or approved by Dymax Corporation will void any effective warranties and may result in damage to the equipment IMPORTANT NOTE DYMAX CORPORATION RESERVES THE RIGHT TO INVALIDATE ANY WARRANTIES EXPRESSED OR IMPLIED DUE TO ANY REPAIRS PERFORMED OR ATTEMPTED ON DYMAX EQUIPMENT WITHOUT WRITTEN AUTHORIZATION FROM DYM...

Страница 15: ...USER GUIDE ACCU CAL 50 LED RADIOMETER 15 ...

Страница 16: ...hat contained in Dymax Europe GmbH s General Terms and Conditions of Sale published on our website Dymax Europe GmbH does not assume any responsibility for test or performance results obtained by users It is the user s responsibility to determine the suitability for the product application and purposes and the suitability for use in the user s intended manufacturing apparatus and methods The user ...